header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Trauma

SQUEAKING INCIDENCE IN HYBRID ALUMINA TOTAL HIP ARTHROPLASTY

European Federation of National Associations of Orthopaedics and Traumatology (EFORT) - 12th Congress



Abstract

Introduction

One of the most common complications of ceramic on ceramic hip replacement is squeaking. The association of Accolade stem and Trident acetabular system has been reported to have squeaking incidence of up to 35,6%. There is doubt if this phenomenon occurs due to: the stem titanium alloy, the V40 femoral neck, the recessed liner of the trident cup or even the mal-seating of the trident insert on the cup.

Objectives

Primary: The purpose of the present study was to determine the incidence of squeaking in association with the use of Exeter stem and Trident ceramic acetabular system. Secondary: Analysis of the correlation of the cup abduction angle and squeaking.

Methods

During the period from March 2004 to December 2008, two surgeons performed 87 total hip arthroplasties in 77 patients with use of a ceramic-on-ceramic bearing (Exeter stem, alumina head, Trident ceramic acetabular system). Seventy six patients (86 THA) were available for review after at least 18 months follow-up. The incidence of squeaking and other noises was analyzed. Cup abduction angle was measured. The Pearson correlation coefficient was used to determine if a correlation existed between the cup abduction angle and squeaking.

Results

The incidence of squeaking was 2,63% (2 patients). Both patients reported a “click” noise in hyperextension of the hip. The mean abduction angle was 44 degress (35–60), and 48 degrees (46 and 50) on the squeaking group. There was no statistically significant difference in the in the mean cup inclination between squeaky and quiet hips.

Conclusion

The incidence of squeaking in association with the use of Exeter stem and Trident ceramic acetabular system was 2,63%. There was no correlation of the cup abduction angle and squeaking.