FINITE ELEMENT ANALYSIS OF CEMENT SHEAR STRESSES IN AUGMENTED TOTAL KNEE REPLACEMENT

British Association for Surgery of the Knee (BASK)



Abstract

Introduction

In complex primary and revision total knee replacement (TKR) the operating surgeon may encounter proximal tibial bone defects. The correct management of such defects is fundamental to both the initial stability and long-term survival of the prosthesis. Block or wedge-shaped metal augments are used to address some such type II unconstrained defects.

Aim

The aim of this finite element (FE) study was to assess the effects of block and wedge-shaped metal augments upon the shear stresses in the cement mantle at the bone-implant interface of an augmented TKR.

Materials and methods

A three-dimensional FE model was constructed from a computer tomography scan of the proximal tibia using SIMPLEWARE v3.2 image processing software. The tibial component of a TKR was implanted with either a block or wedge-shaped metal augment in-situ. The model was axially loaded with a force of 3600N, equating to four times the body weight of a 90kg patient, and the load evenly distributed between the medial and lateral tibial plateaux.

Results

Upon loading of the FE model, shear stresses in the cement-augment interface were found to act towards the centre of the prosthesis. The maximal magnitudes of these ‘reverse’ shear stresses were 3.6MPa with a block-shaped augment and 2.6MPa with a wedge-shaped augment. These values are significantly lower than the reported fatigue limit of cement of 17MPa.

Conclusion

The FE model has demonstrated reduced cement shear stresses with a wedge-shaped rather than block-shaped augment. However, both values of maximal recorded shear stresses are below the fatigue limit of cement. Therefore, either a wedge or block-shaped augments can be used and the choice of augment may be determined by the shape of the defect and the quality of the underlying bone.