Full Access

General Orthopaedics

PREOPERATIVE PREDICTION OF GAP BALANCE BASED ON THE RADIOGRAPHIC FLEXION AND EXTENSION LAXITIES IN ROBOTIC TOTAL KNEE ARTHROPLASTY

The International Society for Computer Assisted Orthopaedic Surgery (CAOS)



Abstract

The preoperative prediction of gap balance after robotic total knee arthroplasty (TKA) is difficult. The purpose of this study was to evaluate the effectiveness of a new method of achieving balanced flexion-extension gaps during robotic TKA.

Fifty one osteoarthritic patients undergoing cruciate retaining TKA using robotic system were included in this prospective study. Preoperative planning was based on the amount of lateral laxity in extension and flexion using varus stress radiograph. After complete milling by the robot and soft tissue balancing, intra-operative extension and flexion gaps were measured using a tensioning device. Knees were subdivided into three groups based on lateral laxities in 0° and 90° of flexion, as follows; the tight extension group (≥ 2mm smaller in extension than flexion laxity), the tight flexion group (≥ 2mm smaller in flexion than extension laxity), and the balanced group (< 2mm difference between laxities). In addition, intra-operative gap balance results were classified as acceptable (0–3mm larger in flexion than in extension), tight (larger in extension than in flexion) or loose (> 3mm larger in flexion than in extension) based on differences between extension and flexion gaps.

During preoperative planning, 34 cases were allocated to the balanced group, 16 to the tight extension group and 1 case was allocated to the tight flexion group. Intra-operative gap balance was acceptable in 46 cases, 4 cases had a tight result, and one case had a loose flexion gap.

We concluded that preoperative planning based on the amount of lateral laxity determined using varus stress radiographs may be useful for predicting intraoperative gap balance and help to achieve precise gap balance during robotic TKA.