header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

THE INNOVATIVE EOS IMAGING SYSTEM FOR THE STUDY OF THA PATIENTS: A PIONEERING EXPERIENCE IN 150 CASES

The International Society for Computer Assisted Orthopaedic Surgery (CAOS)



Abstract

Background

Recent literature points out the potential interest of standing and sitting X-rays for the evaluation of THA patients. The accuracy of the anterior pelvic plane measures is questionable due to the variations in the quality of lateral standing and sitting X-rays. The EOS® (EOS imaging, Paris, France) is an innovative slot-scanning radiograph system allowing the acquisition of radiograph images while the patient is in weightbearing position with less irradiation than standard imagers. This study reports the “functionnal” positions of a 150 THA cohort, including the lateral orientation of the cups.

Methods

The following parameters were measured: sacral slope (SS), pelvic tilt (PT), pelvic incidence (PI) and anterior pelvic plane (APP) sagittal inclination (ASI), frontal inclination (AFI) and planar anteversion (ANT). Irradiation doses were calculated in standing and sitting acquisitions. Variations of sagittal orientation of the cup were measured on lateral standing and sitting images. Descriptive and multivariate analysis were performed for the different parameters studied.

Results

The mean doses for full body were 0,80 mGy ± 0,13 for standing position and 0,94 mGy ± 0,25 for sitting position. The mean value for PI was 55,8° ± 11,4. The mean values standing position were 39,01° ± 9,9 for SS, 17,23° ± 10,2 for PT, and 0,74° ± 8,4 for APP. The mean values were 46,36° ± 9,8 for AFI, 39,49° ± 15,1 for ASI and 22,09° ± 11,1 for ANT. In sitting position, the mean values were 20,87° ± 10,2 for SS, 35,37° ± 13,1 for PT and 21,13° ± 11,2 for APP. The mean values were 56,41° ± 12,3 for AFI, 51,71° ± 14,7 for ASI and 33,45° ± 12,9 for ANT.

Conclusions and Clinical Relevance

Unexpected variations of the anterior pelvic plane can be observed as well as the influence of pelvic incidence on pelvic orientation. The EOS® imaging system provides new informations regarding the pelvis functionnal anatomy in THA patients with potential applications for the study of unstable cases and wear phenomenons.