SIMPLIFIED NAVIGATED DISTAL INTERLOCKING OF INTRAMEDULLARY NAILS: A RADIATION REDUCTION METHOD

The International Society for Computer Assisted Orthopaedic Surgery (CAOS)



Abstract

The common practice for insertion of distal locking screws of intramedullary (IM) nails is a freehand technique under fluoroscopic control. The process is technically demanding, time-consuming and afflicted to considerable radiation exposure to patient and surgical personnel. A new technique is introduced which guides the surgeon by landmarks on the X-ray projection.

18 fresh frozen human below-knee specimens (incl. soft tissue) were used. Each specimen was instrumented with an Expert Tibial Nail (Synthes GmbH, Switzerland) and was mounted on an OR-table. Two distal interlocking techniques were performed in random order using a Siemens ARCADIS C-arm system (Siemens AG, Munich, Germany). The newly developed guided technique, guides the surgeon by visible landmarks projected onto the fluoroscopy image. A computer program plans the drilling trajectory by 2D-3D conversion and provides said guiding landmarks for drilling in real-time. No additional tracking or navigation equipment is needed.

All four distal screws (2 mediolateral, 2 anteroposterior) were placed in each procedure. Operating time, number of taken X-rays and radiation time were recorded per procedure and for each single screw.

8 procedures were performed with the freehand technique and 10 with the guided technique. A 58% reduction in number of fluoroscopy shots per screw was found for the guided technique (7.4±3.4 vs. 17.6±10.3; p < 0.001). Total radiation time was 55% lower for the guided technique (17.1 ± 3.7s vs. 37.9 ± 9.1s) (p = 0.001). Operating time was shorter by 22% in the guided technique (3.2±1.2 min vs. 4.1±2.1 min p = 0.018).

In an experimental setting, the newly developed guided freehand technique has proven to markedly reduce radiation exposure when compared to the conventional freehand technique. The method enhances established clinical workflows and does not require cost intensive add-on devices or extensive training.

A newly developed simple navigated technique has proven to markedly reduce radiation exposure and time for distal locking of intramedullary nails.