DEVELOPMENT OF A PELVIC PHANTOM TO ASSESS ACCURACY OF ANTERIOR PELVIC PLANE DEFINITION AND ACETABULAR ANGLE CALCULATION IN HIP NAVIGATION SYSTEMS

The International Society for Computer Assisted Orthopaedic Surgery (CAOS)



Abstract

There is increasing interest in the use of image free computer assisted surgery (CAS) in total hip arthroplasty (THA). Many of these systems require the registration of the Anterior Pelvic Plane (APP) via the bony landmarks of the anterior superior iliac spines (ASIS) and pubic tubercles (PT) in order to accurately orient the acetabular cup in terms of anteversion and inclination. Given system accuracies are within 1mm and 1° and clinical validation studies have given accuracy by cup position. However, clinical outcomes contain not only system inaccuracies but also variations due to clinical practice. To understand the effects of variation in landmark acquisition on the identification of the acetabular cup orientation, independent bench testing is required. This requires a phantom model that can represent the range of pelvises, male and female, encountered during THA and introduce deliberate known errors to the acquisition to see the effect on anteversion and inclination angles. However, there is a paucity of information in the literature with regards to these specific pelvic dimensions (pelvic width and height). Therefore the aims of this work were to generate the normal expected range of sizes of the APP for both males and females and to use these to manufacture a phantom model that could be used to assess CT free navigation systems.

In the first part of the study 35 human cadavers and 100 pelvic computed tomography (CT) scans were examined.

All cadavers had no gross pelvic abnormalities or previous surgeries. Measurements were carried out with cadavers placed in a supine position. The first author made three sets of measurements using a millimeter ruler. Solid steel pins were used to identify the palpated ASISs and PTs. String was tied between the two ASIS pins and the pelvic width measured. The midpoint of the pubic tubercles was taken to be the midpoint of the pubic symphysis. Pelvic height was measured from the midpoint of the ASIS distance (marked on the string) to the midpoint of the PTs. One hundred pelvic CT scans with no bony abnormalities, previous surgery or metal prosthesis (due to artefacts) were obtained retrospectively from the hospital radiological online system (PACS, Kodak). Mimics software (Mimics12 Materialise, Leuven, Belgium) was used to automatically reconstruct three-dimensional (3D) models using the ‘Bone’ thresholding function. This eliminated any soft tissue from the 3D models. The most anterior ASIS and PT points were then identified on the 3D model surface and measurements of distances made. As the software did not allow identification of points not on the model surface it was not possible to directly obtain the midpoint of the ASIS distance. Therefore to obtain the pelvic height measurements the distance between each ASIS and the ipsilateral and contralateral PTs was also measured. The pelvic height was then calculated using trigonometric functions. The ratio of width to height was calculated (ratio > 1 indicating pelvis width greater than pelvis height). Student's t test was used analyse any differences between male and female pelvic measurements with a p<0.05 being statistically significant.

Using the results from above an aluminium pelvic phantom model was designed and manufactured. It was machined from a billet of marine grade aluminium alloy using a vertical computer numerical controlled (CNC) milling machine. The top surface represented the APP and sides (which represented the acetabuli) were angled to give anteversion and inclination angles of 20° and 45° respectively. Co-ordinates for ASIS and PT points were given based on the 99% prediction intervals from the pelvic data and additional points were milled to give up to a 20 mm error mediolaterally and also in height. Each co-ordinate point was drilled with a 2.0mm diameter ball-nose cutter to a depth of 1.0mm, these holes designed to accommodate the ball-nosed pointer tip to ensure it remained at the same position in space at all orientations of the pointer. Further to this, known errors in height were introduced using accurately manufactured blocks with similar points milled on the surface to fit a ball-nosed pointer. These blocks could be secured to the top surface of the model using screws. A Perspex base unit with tracker attachments was made to hold the phantom and provide the reference frame. A further support that enables the phantom to also be used in the “lateral” position was manufactured.

For the assessment of pelvic size there were 66 females and 69 males, mean age 62.3 years (range from 20 to 99 years). The mean width was 238 mm (SD 20 mm) and mean height was 93 mm (SD 11 mm) with a mean ratio of 2.6 (SD 0.3). There were no statistically significant differences in mean between males and females (p>0.4 in all cases). From this data set the range of APP sizes required to cover 99% of population (width 186 to 290 mm and height 66 to 120 mm) and therefore the measurements for the model were generated. The manufactured model can be used to give the range of pelvis sizes from 170mm to 290mm in width and 60mm to 120mm in height and also to add up to 20 mm of error in palpation of each of the ASISs and PT.

This study generated APP sizes to cover 99% of the general population over a wide age range. It illustrated that a single pelvic model would fit both sexes. The model allows the determination of the effects of changes of the pelvic dimensions may have on the acetabular orientation measured on an image free CAS system including the assessment of point acquisition and deliberate errors. The model has been successfully used in preliminary testing and can be used to assess any CT free system.