RESULTS OF 37 COMPUTER-ASSISTED DOUBLE LEVEL OSTEOTOMIES FOR SEVERE GENU VARUM DEFORMITY: A 1 TO 9 YEARS FOLLOW-UP

The International Society for Computer Assisted Orthopaedic Surgery (CAOS)



Abstract

Double level osteotomy (DLO) for severe genu varum is not a common technique. We performed our first computer-assisted double level osteotomy (CADLO) in March 2001 and we published our preliminary results in 2005 and 2007. The rationale to perform this procedure is to avoid oblique joint line in order to have less difficulty in case of revision to a total knee arthroplasty (TKA). The goal of this paper is to present the results of 37 cases operated on between August 2001 and January 2010.

The series was composed of 35 patients (two bilateral), nine females and 26 males, aged from 39 to 64 years old (mean age: 50.5 +/− 7.5). We operated on 20 right knees and 17 left ones. The mean BMI was 29.3 +/− 4.3 for a mean height of 1.71 m and a mean weight of 85.8 kg. The functional status was evaluated according to the LYSHÖLM and TEGNER score. The mean score was of 42.4 +/− 8.9 points (22–69). According to modified AHLBÄCK criteria we operated on seven stage 2, 22 stage 3, five stage 4 and two stage 5. We measured HKA (Hip-Knee-Ankle) angle using RAMADIER's protocol and we also measured the femoral mechanical axis (FMA) and the tibial mechanical axis (TMA) to pose the right indication. These measures were respectively: 168° +/− 3.4° (159°–172°), 87.5° +/− 2.1 (83°–91°) for the FMA and 83.7° +/− 2.6° (78°–88°) for the TMA.

The inclusion criteria were a patient younger than 65 years old with a severe varus deformity (more than 8° − HKA angle ≤ to 172°) and a FMA at 91° or less. All the osteotomies were navigated using the ORTHOPILOT® device (B-BRAUN-AESCULAP, TUTTLINGEN, GERMANY). The procedure was performed as follows: after inserting the rigid-bodies and calibrating the lower leg, we did first the femoral closing wedge osteotomy (from 4 to 7 mm) which was fixed by a an AO T-Plate, and secondly, after checking the residual varus, the high tibial opening wedge osteotomy using a BIOSORB® wedge (Tricalcium phosphate) and a plate (AO T-plate or C-plate). The goals of the osteotomy were to achieve an HKA angle of 182° +/− 2° and a TMA angle of 90° +/− 2°.

The functional results were evaluated using the LYSHÖLM-TEGNER score and the KOOS score. The patients answered the questionnaire at revision or by phone, and the radiological results were assessed by plain radiographs and standing long leg X-Rays between three and six months postoperatively.

We had no complication in this series but one case of recurrence of the deformity related to an impaction of the femoral osteotomy on the medial side. Two patients were lost to follow-up after removing of the plates (24 months) but were included in the results because the file was complete at that date. All the patients were assessed at a mean follow-up of 43 +/− 27 months (12–108). The mean LYSHÖLM-TEGNER score was 78.7 +/− 7.5 points (59–91) and the mean KOOS score was 94.9 +/− 3.3 points (89–100). Thirty-five patients were satisfied (18) or very satisfied (17) of the result. Only two were poorly satisfied. Regarding the radiological results, if we exclude the patient who had a loss of correction, the goals were reached in 32 cases (89%) for the HKA angle and in 31 cases (86%) for the TMA with only one case at 93°. The mean angles were: 181.97° +/− 1,89° (177°–185°) for HKA, 89.86° +/− 1,85° (85°–93°) for TMA and 93.05° +/− 2.3° (89°–99°) for FMA. At that mid-term follow-up no patient had revision to a total knee arthroplasty.

DLO is a very demanding technique. Navigation can improve the accuracy of the correction compared to non computer-assisted osteotomies. The functional results are satisfying and the satisfaction of the patients is very high. Despite the difficulty of the procedure, complications are, in our hands, very rare. We recommend DLO for severe genu varum deformity in young patients to avoid oblique joint line, which will be difficult to revise to TKA.