header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

A BIOMECHANICAL EVALUATION OF SHAPE MEMORY STAPLES

Australian Orthopaedic Association and New Zealand Orthopaedic Association (AOA/NZOA) - Combined Annual Scientific Meeting



Abstract

Shape memory staples have several uses in both hand and foot and ankle surgery. There is relatively little data available regarding the biomechanical properties of staples, in terms of both the compression achieved and potential decay of mechanical advantage with time. An understanding of these properties is therefore important for the surgeon.

Two blocks of synthetic polyurethane mimicking properties of cancellous bone were fixed in jigs to both the actuator and 6 degree-of-freedom load cell of an MTS servohydraulic testing machine. With the displacement between the blocks held constant the peak value and subsequent decay in compressive force applied by both the smooth and barbed version of the nitinol OSStaple (Biomedical Enterprises), Easyclip (LMT), Herbert Bone Screws (Martin) and the Headless Compression Screw (Synthes) was measured. Nitinol staples were energised once only. A second experiment was conducted to assess the effects of repeated energisation on these parameters.

The Easyclip staples achieved a mean peak force of 5.2N, whilst the smooth and barbed OSStaples achieved values of 9.3N and 5.7N, respectively. The Herbert screws achieved a mean peak force of 9N and the headless compression screws 23.9N. The mean peak force achieved with 2 Easyclip staples in parallel was 8.1N. Following the application of a single energisation the OSStaples exhibited a significant reduction in compressive load, losing up to approximately 70% of the peak value attained. The repeated energisation of these nitinol staples produced progressive increases in both peak and trough loads, the positive effects exhibited a plateau with time.

Performance of both OSStaples was comparable to the Herbert screw with regard to reduction load applied across a simulated fracture plane. The maximum load applied by the OSStaples diminished with time. Staples provide fixation without violating the fracture plane which has the potential to offer some benefits from a healing perspective.