Abstract

Introduction

Osteotomies around the knee are traditionally templated on 2D plain X-rays. Results are often inaccurate and inconsistent and multiplanar ostetomies are hard to perform. The aim of this study is to evaluate the feasibility and accuracy of virtual three-dimensional CT-based planning and correct execution of osteotomies around the knee with the aid of patient specific surgical guides and locking plates.

Methods

Eight consecutive patients with significant malalignment of the lower limb were included in the study. Pre-operative CT scans of the affected limb and the normal contra-lateral side were obtained and 3D models of the patient's anatomy were created, using dedicated software. The healthy contralateral limb was mirrored and geometrically matched to the distal femur or proximal tibia of the healthy side. A virtual opening wedge correction of the affected bone was used to match the geometry of the healthy contralateral bone. Standard lower limb axes measurements confirmed correction of the alignment. Based on the virtual plan, surgical guides were designed to perform the planar osteotomy and achieve the planned wedge opening and hinge axis orientation. The osteotomy was fixed with locking plates and screws. Post-operative assessment included planar X-rays, CT-scan and full leg standing X-rays.

Results

One three-planar, three bi-planar and four single-plane osteotomies were performed. Maximum weightbearing mechanical femoro-tibial coronal malalignment varied between 7° varus and 14° valgus (mean 7.6°, SD 3.1). Corrective angles varied from 7°-15°(coronal), 0°–13°(sagittal) and 0°–23°(horizontal). The maximum deviation between the planned pre-operative wedge angle and the executed post-operative wedge angle was 1° in the coronal, sagittal and horizontal plane. The desired mechanical femorotibial axis on full-leg standing X-rays was achieved in 6 patients. Two patients were undercorrected by 1° and 2° respectively.

Conclusion

3D planning and guided correction of multi-planar deformity of femur or tibia is a feasible and accurate novel technique.