header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

TOTAL KNEE ARTHROPLASTY REVISION RATES IN MOBILE- VERSUS FIXED-BEARING IMPLANTS: A NEW ZEALAND JOINT REGISTRY STUDY

The New Zealand Orthopaedic Association (NZOA) Annual Scientific Meeting 2023, Nelson, New Zealand, 5–7 November 2023.



Abstract

Introduction

The mobile-bearing (MB) total knee arthroplasty (TKA) design was introduced with the aim of reducing polyethylene wear and component loosening seen in the fixed-bearing (FB) design. A recent joint registry study has revealed increased risk for all-cause revision, but not revision for infection, in MB-TKA. We used the New Zealand Joint Registry (NZJR) to compare all-cause revision rates, and revision rates for aseptic loosening of MB-TKA compared with fixed bearing (FB) TKA.

Methods

All patients who underwent a primary TKA registered in the NZJR between the 1st January 1999 to 31st December 2021 were identified. Analysis compared MB to FB designs, with sub analysis of implants from a single company. We identified 135,707 primary TKAs, with 104,074 (76.7%) FB-TKAs and 31,633 (23.3%) MB-TKAs recorded. We examined all-cause revision rates, reasons for revision and performed survival analyses.

Results

For all-comers, MB-TKA had an all-cause revision rate of 0.43/100-component-years (OCY) compared with 0.42/OCY for FB-TKA (p=0.09). The all-cause revision rate was higher for those age < 65 years (MB TKA 0.60/OCY vs. FB-TKA 0.59/OCY) compared to those > 65 years at time of primary TKA (MB-TKA 0.29/OCY vs. FB-TKA 0.32/OCY), however there was no statistically significant difference between implant design in either age group (p=0.16 and p=0.64; respectively). Similarly, there was no difference in revision rates for aseptic loosening between implant designs. Kaplan-Meier survival analysis demonstrates no statistically significant difference in revision-free survival of implants, with both MB-TKA and FB-TKA demonstrating ∼93% revision free survival at 23 years.

Conclusions

Both FB- and MB-TKA demonstrated excellent survivorship, with no significant difference in all-cause revision rates or revision for aseptic loosening between implant designs.