AN IN VITRO STUDY EXAMINING WIRE TENSION IN FINE WIRE CIRCULAR EXTERNAL FIXATION CONSTRUCTS

The British Limb Reconstruction Society (BLRS) Annual Meeting 2024, Glasgow, Scotland, 14–15 March 2024.



Abstract

Introduction

Ilizarov fixators are reliant on tensioned fine wires for stability. The tension in the wires is generated using specific tensioning devices. Loss of wire tension over time may lead of loss a stability and complications. A series of in vitro experiments were undertaken to explore wire tensioner accuracy, the impact of fixation bolt torque and initial tension on loss of tension in ilizarov constructs under static and dynamic loads.

Materials & Methods

Medical grade materials were applied to a synthetic bone analogue using surgical instruments in all experiments. Bolt torque was fixed at 6, 10 or 14 Nm using a torque limiting wrench. Wire tension was assessed using a strain measurement bridge. Wires were tensioned to 90, 110 and 130kg as measured by a commercial dynamometric tensioner. Static and dynamic testing was undertaken using an instron testing machine. Cyclical loads from 50–750N were applied for 5000 cycles.

Results

Actual wire tension was approximately 15% less than indicated by the tensioner device. Using fixation bolt torques of 10Nm and 14Nm achieved final wire tensions of around 60% and 80% of that applied at 90 and 130kg of applied tension. Static load testing demonstrated self stiffening to similar levels in all pre-tensions. Dynamic testing demonstrated significant loss of tension, most of which occured in the first 3 cycles, inversely proportional to the tension initially applied.

Conclusions

These experiments provides insight into the effect of initially applied wire tension on Illizarov mechanical performance. It is important surgeons understand how the different ways that these devices are applied affects mechanical performance. Further research examining what factors affect performance across different manufacturers equipment would therefore be relevant, alongside the development of novel fixation methods to reduce wire slippage and the further development of equipment for clinical use.