header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Trauma

CEPHALOMEDULLARY NAILING FOR SUBTROCHANTERIC FEMORAL SHAFT FRACTURES: A COMPARISON OF SINGLE LAG SCREW WITH A DUAL LAG SCREW DEVICE

The Scottish Committee for Orthopaedics and Trauma (SCOT) Meeting, Crieff, Scotland, 19 January 2024.



Abstract

This study compares outcomes of fixation of subtrochanteric femoral fractures using a single lag screw (Gamma3 nail, GN) with a dual lag screw device (InterTAN nail, IN). The primary outcome was mechanical failure, defined as lag screw cut-out, back-out, nail breakage or peri-implant fracture. Technical factors associated with mechanical failure were also identified.

All adult patients (>18yrs) with a subtrochanteric femoral fracture treated in a single centre were retrospectively identified using electronic records. Included patients underwent surgical fixation using either a long GN (2010–2017) or IN (2017–2022). Cox regression analysis was used to determine the risk of mechanical failure and technical predictors of failure.

The study included 587 patients, 336 in the GN group (median age 82yrs, 73% female) and 251 in the IN group (median age 82yrs, 71% female). The IN group exhibited a higher prevalence of osteoporosis (p=0.002) and CKD□3 (p=0.007). There were no other baseline differences between groups. The risk of any mechanical failure was increased two-fold in the GN group (HR 2.51, p=0.020). Mechanical failure comprising screw cut-out (p=0.040), back-out (p=0.040) and nail breakage (p=0.51) was only observed in the GN group. The risk of peri-implant fracture was similar between the groups (HR 1.10, p=0.84). Technical predictors of mechanical included varus >5° for cut-out (HR 15.61, p=0.016), TAD>25mm for back-out (HR 9.41, p=0.020) and shortening >1cm for peri-implant fracture (HR 6.50, p=<0.001).

Dual lag screw designs may reduce the risk of mechanical complications for patients with subtrochanteric femoral fractures.


Email: