DETERMINING THE ACCURACY OF THE SMART BRACE USING MULTI-FREQUENCY BIOIMPEDANCE ANALYSIS TO MEASURE KNEE SWELLING

The European Orthopaedic Research Society (EORS) 31st Annual Meeting, Porto, Portugal, 27–29 September 2023. Part 2 of 2.



Abstract

Knee swelling is common after injury or surgery, resulting in pain, restricted range of movement and limited mobility. Accurately measuring knee swelling is critical to assess recovery. However, current measurement methods are either unreliable or expensive [1,2]. Therefore, a new measurement method is developed. This wearable (the ‘smart brace’) has shown the ability to distinguish a swollen knee from a not swollen knee using multi-frequency-bio impedance analysis (MF-BIA) [3].

This study aimed to determine the accuracy of this smart brace. The study involved 25 usable measurements on patients treated for unilateral knee osteoartritis with a 5mL injection of Lidocaïne + DepoMedrol (1:4). MF-BIA measurements were taken before and after the injection, both on the treated and untreated knee. The smart brace accurately measured the effect of the injection by a decrease in resistance of up to 2.6% at 100kHz (p<0.01), where commonly used gel electrodes were unable to measure the relative difference. Remarkably, both the smart brace and gel electrodes showed a time component in the MF-BIA measurements.

To further investigate this time component, 10 participants were asked to lie down for 30 minutes, with measurements taken every 3 minutes using both gel electrodes and the smart brace on both legs. The relative change between each time step was calculated to determine changes over time. The results showed presence of a physiological aspect (settling of knee fluids), and for the brace also a mechanical aspect (skin-electrode interface) [4]. The mechanical aspect mainly interfered with reactance values.

Overall, the smart brace is a feasible method for quantitatively measuring knee swelling as a relative change over time. However, the skin-electrode interface should be improved for reliable measurements at different moments in time. The findings suggest that the smart brace could be a promising tool for monitoring knee swelling during rehabilitation.