PRE-CLINICAL EVALUATION OF CUSTOM-MADE KNEE IMPLANTS USING FINITE ELEMENT ANALYSIS

The International Combined Orthopaedic Research Societies (ICORS), World Congress of Orthopaedic Research, Edinburgh, Scotland, 7–9 September 2022. Part 2 of 3.



Abstract

In severe cases of total knee arthroplasty which cannot be treated with off-the-shelf implants anymore custom-made knee implants may serve as one of the few remaining options to restore joint function or to prevent limb amputation. Custom-made implants are specifically designed and manufactured for one individual patient in a single-unit production, in which the surgeon is responsible for the implant design characteristics in consultation with the corresponding engineer.

The mechanical performance of these custom-made implants is challenging to evaluate due to the unique design characteristics and the limited time until which the implant is needed. Nevertheless, the custom-made implant must comply with clinical and regulatory requirements. The design of custom-made implants is often based on a underlying reference implant with available biomechanical test results and well-known clinical performance. To support surgeons and engineers in their decision whether a specific implant design is suitable, a method is proposed to evaluate its mechanical performance.

The method uses finite element analysis (FEA) and comprises six steps: (1) Identification of the main potential failure mechanism and its corresponding FEA quantity of interest. (2) Reproduction of the biomechanical test of the reference implant via FEA. (3) Identification of the maximum value of the corresponding FEA quantity of interest at the required load level. (4) Definition of this value as the acceptance criteria for the FEA of the custom-made implant. (5) Reproduction of the biomechanical test with the custom-made implant via FEA. (6) Conclusion whether the acceptance criteria is fulfilled or not.

The method was applied to two exemplary cases of custom-made knee implants. The FEA acceptance criteria derived from the reference implants were fulfilled in both custom-made implants. Subsequent biomechanical tests verified the FEA results.

This study suggests and applies a non-destructive and efficient method for pre-clinical testing of a single-unit custom-made knee implant to evaluate whether the design is mechanically suitable.