header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

Research

VALIDATION OF A NOVEL DEVICE FOR DIGITALLY ENHANCED HANDS-ON SURGICAL TRAINING OF INTRAMEDULLARY NAIL DISTAL INTERLOCKING

The International Combined Orthopaedic Research Societies (ICORS), World Congress of Orthopaedic Research, Edinburgh, Scotland, 7–9 September 2022. Part 2 of 3.



Abstract

Freehand distal interlocking of intramedullary nails is technical demanding and prone to handling issues. It requires the surgeon to precisely place a screw through the nail under x-ray. If not performed accurately it can be a time consuming and radiation expensive procedure. The aims of this study were to assess construct and face validity of a new training device for distal interlocking of intramedullary nails.

53 participants (29 novices and 24 experts) were included. Construct validity was evaluated by comparing simulator metrics (number of x-rays, nail hole roundness, drill tip position and accuracy of the drilled hole) between experts and novices. Face validity was evaluated by means of a questionnaire concerning training potential and quality of simulated reality using a 7-point Likert scale (range 1-7).

Mean realism of the training device was rated 6.3 (range 4-7) and mean training potential as well as need for distal interlocking training was rated 6.5 (range 5-7) with no significant differences between experts and novices, p≥0.236. All participants stated that the simulator is useful for procedural training of distal nail interlocking, 96% would like to have it at their institution and 98% would recommend it to their colleagues. Total number of x-rays were significantly higher for novices (20.9±6.4 vs. 15.5±5.3), p=0.003. Successful task completion (hit the virtual nail hole with the drill) was significantly higher in experts (p=0.04; novices hit: n=12; 44,4%; experts hit: n=19; 83%).

The evaluated training device for distal interlocking of intramedullary nails yielded high scores in terms of training capability and realism. Furthermore, construct validity was established as it reliably discriminates between experts and novices. Participants see a high further training potential as the system may be easily adapted to other surgical task requiring screw or pin position with the help of x-rays.


Email: