Knee

A RANDOMIZED CONTROLLED TRIAL OF KINEMATICALLY AND MECHANICALLY ALIGNED TOTAL KNEE ARTHROPLASTY: 12-YEAR FOLLOW-UP

The British Association for Surgery of the Knee (BASK) May 2023 Meeting, London, England, 16–17 May 2023.



Abstract

Abstract

Introduction

The optimal alignment technique for total knee replacement (TKR) remains controversial. We previously reported six-month and two-year results of a randomized controlled trial comparing kinematically (KA) versus mechanically (MA) aligned TKR. In the present study, we report 12-year results from this trial.

Methods

The original cohort included 88 TKRs (44 KA using Shape Match patient-specific guides and 44 MA using conventional instrumentation), performed from 2008 to 2009. After IRB approval, the health record of the original 88 patients were queried. Revisions, re-operations, and complications were recorded. The non-deceased patients were contacted via phone. Reoperation and complications were documented via the patient's history. Further, a battery of patient-reported outcome measures (including patient satisfaction, WOMAC, Oxford, KOOS Jr, Forgotten Joint Score, and M-SANE) were obtained.

Results

Of the original 88 patients in the study, 15 patients had a least one reoperation (17%). Patella problems were the most common cause of reoperation accounting for 5/8 reoperations in the KA group versus 3/7 in the MA group. There was no statistically significant difference between the two alignment methods in terms of major and minor complications or reoperations. At the 12-year follow-up, 26 patients died leaving 62 patients for follow-up. Of these, 48 patients (77%) were successfully contacted. The kinematically aligned total knees self-reported better satisfaction (96% versus 82%), but no difference in other patient-reported outcome measures compared to mechanically aligned TKRs.

Conclusion

KA TKR demonstrates excellent mid to long-term results compared to MA TKR with similar reoperations, complications, and patient-reported outcome measures.