Abstract

Modular dual mobility (DM) articulations are increasingly utilized during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS) magnetic resonance imaging (MRI) abnormalities with serum metal ion levels in patients with DM articulations.

All patients with an asymptomatic, primary THA and DM articulation with >2-year follow-up underwent MARS-MRI of the operative hip. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.

Forty-five patients (50 hips) with a modular DM articulation were included with average follow-up of 3.7±1.2 years. Two patients (4.4%) had abnormal periprosthetic fluid collections on MARS-MRI with cobalt levels >3.0 μg/L. Four patients (8.9%) had MARS-MRI findings consistent with greater trochanteric bursitis, all with cobalt levels < 1.0 μg/L. A seventh patient had a periprosthetic fluid collection with normal ion levels. Of the 38 patients without MARS-MRI abnormalities, 37 (97.4%) had cobalt levels <1.0 μg/L, while one (2.6%) had a cobalt level of 1.4 μg/L. One patient (2.2%) had a chromium level >3.0 μg/L and a periprosthetic fluid collection. Of the 41 patients with titanium levels, five (12.2%) had titanium levels >5.0 μg/L without associated MARS-MRI abnormalities.

Periprosthetic fluid collections associated with elevated serum cobalt levels in patients with asymptomatic dual mobility articulations occur infrequently (4.4%), but further assessment of these patients is necessary.

Level of Evidence: Level IV