header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

THE FITBONE LEARNING CURVE: WHEN DOES IT END?

The British Limb Reconstruction Society (BLRS) Annual Meeting 2023, Belfast, Northern Ireland, 23–24 March 2023.



Abstract

Introduction

The FitBone lengthening nail (Orthofix UK) is an intramedullary device licensed for the lengthening of long bones in adults in the UK. It contains a motor powered by electricity transmitted via an induction coil placed underneath the skin. It was developed in Germany two decades ago but uptake in the UK has only started more recently. The aim of this study was to review the first cohort of FitBone lengthening nails in a unit with significant experience of other lengthening nails (including PRECICE and Stryde).

Materials & Methods

Demographic, clinical and radiological data was prospectively collected on all FitBone cases starting in February 2022. Accuracy of lengthening rate, patient satisfaction and implant issues were all considered. Complications and learning points were recorded and discussed by the multidisciplinary team involved in the patients care.

Results

Eleven lengthening nails were inserted between February and November 2022 (6 right femurs, 5 left femurs). The average patient age was 31 (16–57) with 4 females and 7 males. The average lengthening achieved was 44mm (13– 70) over an average of 59 days (35 to 104). Significant technical issues were encountered in this cohort of patients including slow opening up at osteotomy site (3 requiring speeding up of programme), early consolidation (one requiring re-do osteotomy) and backing out of locking screws (3 out of 11 nails). There were also patient use concerns with difficulty using the motor and the inability to reverse the lengthening without an additional component to the motor.

Conclusions

We present the first UK cohort of patients with femoral lengthening using the FitBone implant and device. We highlight the technical and patient issues encountered during this learning curve and propose solutions to avoid these pitfalls.