header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

CLINICAL TRIAL OF A NEW DEVICE FOR REAL-TIME MUSCLE PRESSURE MEASUREMENTS IN PATIENTS WITH AN UPPER OR LOWER EXTREMITY FRACTURE AT RISK FOR ACUTE COMPARTMENT SYNDROME

The Canadian Orthopaedic Association (COA) and Canadian Orthopaedic Research Society (CORS) Annual General Meeting, Quebec City, Quebec, Canada, 8–11 June 2022. Part 2 of 2.



Abstract

Acute Compartment Syndrome (ACS) is an orthopaedic emergency that can develop after a wide array of etiologies. In this pilot study the MY01 device was used to assess its ease of use and its ability to continuously reflect the intracompartmental pressure (ICP) and transmit this data to a mobile device in real time. This preliminary data is from the lead site which is presently expanding data collection to five other sites as part of a multi-center study.

Patients with long bone trauma of the lower or upper extremity posing a possibility of developing compartment syndrome were enrolled in the study. Informed consent was obtained from the patients. A Health Canada licensed continuous compartmental pressure monitor (MY01) was used to measure ICP. The device was inserted in the compartment that was deemed most likely to develop ACS and ICP was continuously measured for up to 18 hours. Fractures were classified according to the AO/OTA classification. Patient clinical signs and pain levels were recorded by healthcare staff during routine in-patient monitoring and were compared to the ICP from the device. Important treatment information was pulled from the patient's chart to help correlate all of the patient's data and symptoms.

The study period was conducted from November 2020 through December 2021. Twenty-six patients were enrolled. There were 17 males, and nine females. The mean age was 38 years (range, 17–76). Seventeen patients received the device post-operatively and nine received it pre-operatively. Preliminary results show that post-operative ICPs tend to be significantly higher than pre-operative ICPs but tend to trend downwards very quickly. The trend in this measurement appears to be more significant than absolute numbers which is a real change from the previous literature. One patient pre-operatively illustrated a steep trend upwards with minimal clinical symptoms but required compartment release at the time of surgery that exhibited no muscle necrosis. The trend in this patient was very steep and, as predicted, predated the clinical findings of compartment syndrome. This trend allows an early warning signal of the absolute pressure, to come, in the compartment that is being assessed by the device.

Preliminary results suggest that this device is reliable and relatively easy to use within our institutions. In addition it suggests that intracompartmental pressures can be higher immediately post-op but lower rapidly when the patient does not develop ACS. These results are in line with current literature of the difference between pre and post-operative baselines and thresholds of ICP, but are much more striking, as continuous measurements have not been part of the data set in most of past studies.

Further elucidation of the pressure thresholds and profiles are currently being studied in the ongoing larger multicenter study and will add to our understanding of the critical values. This data, plus the added value of continuous trends in the pressure, upwards or downwards, will aid in preventing muscle necrosis during our management of these difficult long bone fractures.