header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

LUMIC® ENDOPROSTHESIS FOR PELVIC RECONSTRUCTION: A CANADIAN EXPERIENCE

The Canadian Orthopaedic Association (COA) and Canadian Orthopaedic Research Society (CORS) Annual General Meeting, Quebec City, Quebec, Canada, 8–11 June 2022. Part 2 of 2.



Abstract

The reconstruction of peri-acetabular defects after severe bone loss or pelvic resection for tumor is among the most challenging surgical intervention. The Lumic® prosthesis (Implantcast, Buxtehude, Germany) was first introduced in 2008 in an effort to reduce the mechanical complications encountered with the classic peri-acetabular reconstruction techniques and to improve functional outcomes. Few have evaluated the results associated with the use of this recent implant.

A retrospective study from five Orthopedic Oncology Canadian centers was conducted. Every patient in whom a Lumic® endoprosthesis was used for reconstruction after peri-acetabular resection or severe bone loss with a minimal follow-up of three months was included. The charts were reviewed and data concerning patients’ demographics, peri-operative characteristics and post-operative complications was collected. Surgical and functional outcomes were also assessed.

Sixteen patients, 11 males and five females, were included and were followed for 28 months [3 – 60]. Mean age was 55 [17-86], and mean BMI reached 28 [19.6 – 44]. Twelve patients (75%) had a Lumic® after a resection of a primary sarcoma, two following pelvic metastasis, one for a benign tumor and one after a comminuted acetabular fracture with bone loss. Twelve patients (75%) had their surgery performed in one stage whereas four had a planned two-stage procedure. Mean surgical time was 555 minutes [173-1230] and blood loss averaged 2100 mL [500-5000]. MSTS score mean was 60.3 preoperatively [37.1 – 97] and 54.3 postoperatively [17.1-88.6]. Five patients (31.3%) had a cemented Lumic® stem. All patients got the dual mobility bearing, and 10 patients (62.5%) had the largest acetabular cup implanted (60 mm). In seven of these 10 patients the silver coated implant was used to minimize risk of infection. Five patients (31.3%) underwent capsular reconstruction using a synthetic fabric aiming to reduce the dislocation risk. Five patients had per-operative complications (31.3%), four were minor and one was serious (comminuted iliac bone fracture requiring internal fixation). Four patients dislocated within a month post-operatively and one additional patient sustained a dislocation one year post-operatively. Eight patients (50%) had a post-operative surgical site infection. All four patients who had a two-stage surgery had an infection. Ten patients (62.5%) needed a reoperation (two for fabric insertion, five for wash-outs, and three for implant exchange/removal). One patient (6.3%) had a septic loosening three years after surgery. At the time of data collection, 13 patients (81.3%) were alive with nine free of disease. Silver coating was not found to reduce infection risk (p=0.2) and capsuloplasty did not prevent dislocation (p=1).

These results are comparable to the sparse data published. Lumic® endoprosthesis is therefore shown to provide good functional outcomes and low rates of loosening on short to medium term follow-up. Infection and dislocation are common complications but we were unable to show benefits of capsuloplasty and of the use of silver coated implants. Larger series and longer follow-ups are needed.


Email: