OUTCOMES FOLLOWING SURGICAL MANAGEMENT OF MULTI-LIGAMENTOUS KNEE INJURIES USING ADJUSTABLE SUSPENSORY FIXATION

The Canadian Orthopaedic Association (COA) and Canadian Orthopaedic Research Society (CORS) Annual General Meeting, Quebec City, Quebec, Canada, 8–11 June 2022. Part 2 of 2.



Abstract

Ligament reconstruction following multi-ligamentous knee injuries involves graft fixation in bone tunnels using interference screws (IS) or cortical suspensory systems. Risks of IS fixation include graft laceration, cortical fractures, prominent hardware, and inability to adjust tensioning once secured. Closed loop suspensory (CLS) fixation offers an alternative with fewer graft failures and improved graft-to-tunnel incorporation. However, graft tensioning cannot be modified to accommodate errors in tunnel length evaluation. Adjustable loop suspensory (ALS) devices (i.e., Smith & Nephew Ultrabutton) address these concerns and also offer the ability to sequentially tighten each graft, as needed. However, ALS devices may lead to increased graft displacement compared to CLS devices. Therefore, this study aims to report outcomes in a large clinical cohort of patients using both IS and CLS fixation.

A retrospective review of radiographic, clinical, and patient-reported outcomes following ligament reconstruction from a Level 1 trauma centre was completed. Eligible patients were identified via electronic medical records using ICD-10 codes. Inclusion criteria were patients 18 years or older undergoing ACL, PCL, MCL, and/or LCL reconstruction between January 2018 and 2020 using IS and/or CLS fixation, with a minimum of six-month post-operative follow-up. Exclusion criteria were follow-up less than six months, incomplete radiographic imaging, and age less than 18 years. Knee dislocations (KD) were classified using the Schenck Classification. The primary outcome measure was implant removal rate. Secondary outcomes were revision surgery rate, deep infection rate, radiographic fixation failure rate, radiographic malposition, Lysholm and Tegner scores, clinical graft failure, and radiographic graft failure. Radiographic malposition was defined as implants over 5 mm off bone or intraosseous deployment of the suspensory fixation device. Clinical graft failure was defined as a grade II or greater Lachman, posterior drawer, varus opening at 20° of knee flexion, and/or valgus opening at 20° of knee flexion. Radiographic failure was defined when over 5 mm, 3.2 mm, and/or 2.7 mm of side-to-side difference occurred using PCL gravity stress views, valgus stress views, and/or varus stress views, respectively. Descriptive statistics were used.

Sixty-three consecutive patients (mean age = 41 years, range = 19-58) were included. A total of 266 CLS fixation with Ultrabuttons and 135 IS were used. Mean follow-up duration was 383 days. Most injuries were KD type II and III. Graft revision surgery rate was 1.5%. Intraosseous deployment occurred in 6.2% and 17% had implants secured in soft tissue, rather than on bone. However, the implant removal rate was only 6.2%. Radiographic PCL gravity stress views demonstrated an average of 1.2 mm of side-to-side difference with 6.2% meeting criteria for radiographic failure. A single patient met radiographic failure criteria for collateral grafts. Mean Lysholm and Tegner scores were 87.3 and 4.4, respectively, with follow-up beyond one year.

Both IS and CLS fixation demonstrate an extremely low revision surgery rate, a high rate of implant retention, excellent radiographic stability, and satisfactory patient-reported outcome scores.

Incorrect implant deployment was seen in a total of 17% of patients, yet none required implant removal. A single patient required graft revision due to implant failure.