Abstract
Abstract
Background
Multiple devices can stabilise the MTP joint for arthrodesis. The ideal implant should be easy to use, provide reproducible and high quality results, and ideally enable early rehabilitation to enable faster return to function, whilst lessening soft tissue irritation. We prospectively evaluated the combination of the IO-Fix (Extremity Medical, NJ, USA) device which consists of an intra-osseous post and lag screw that offers these features with full bearing of weight after surgery.
Methods
67 feet in 65 patients were treated over 31 months. After excluding patients lost to follow-up, undergoing revision arthrodesis, or concomitant first ray procedures, there were 54 feet in 52 patients available with a minimum 12 month follow-up with clinical and radiographic outcomes. All patients were treated using a similar operative technique with immediate bearing of weight in a rigid soled shoe.
Results
The mean MOXFQ score improved from 46.4 (range 18 – 64) before surgery to 30.2 (range 0 – 54) at 6 months after surgery (p=0.02), and 18.4 (range 0 – 36) (p< 0.001) at latest follow-up. Arthrodesis across the MTP joint was achieved in 52 feet (96%), at a mean of 61 days (range 39–201). Non-union was observed in two feet; superficial wound infections in two feet; and metalwork impingement in three feet.
Conclusions
In the largest reported series to date, the IO-Fix device achieved a union rate of 96% across the MTP joint when coupled with immediate bearing of weight. Significant improvements were seen in patient reported outcomes with low complication rates.