GPS NAVIGATION SYSTEM ALLOWS THE SURGEON TO PREPARE THE IMPLANT SITE AS PLANNED ON PREOPERATIVE SOFTWARE IN REVERSE SHOULDER ARTHROPLASTY

International Society for Technology in Arthroplasty (ISTA) meeting, Emerging Technologies in Arthroplasty (ETA), held online, 15 May 2021.



Abstract

Introduction

Reverse shoulder Arthroplasty is a successful treatment for gleno-humeral osteoarthritis. However, components loosening and painful prostheses, related to components wrong positioning, are still a problem for those patients who underwent this kind of surgery. Several new technology has been developed the improve the implant positioning. CT-based intraoperative navigation system is a suitable technology that allow the surgeon to prepare the implant site exactly as planned with preoperative software.

Method

Thirty reverse shoulder prostheses were performed at Modena Polyclinic using GPS CT-based intraoperative navigation system (Exactech, Gainsville, Florida). Walch classification was used to assess glenoid type. Planned version and inclination of the glenoid component, planned seating, final version and inclination of the reamer were recorded. Intraoperative and perioperative complication were recorded. Planned positioning was conducted aiming to the maximum seating, avoiding retroversion >10° and superior inclination.

Results

Eight patients were male, 22 were female. Mean age was 75 years old (range 58–87). 4 glenoid were type B3, four were B2, 10 cases were B1, 12 case were A1/A2. Posterior or superior augment was used in 15 cases. Mean planned seating was 93%.

Mean preoperative version was -7.5±6.9°; Mean planned version was -2±2.8°; Mean intraoperative measured version was -1.9±2.8°; no statistical difference was found between planned and intraoperative version (p=0.16). Mean preoperative inclination was 1.8±6.°; Mean planned inclination was -2.2±2.4°; Mean intraoperative measured inclination was -2.1.9±2.3°; no statistical difference was found between planned and intraoperative version or inclination (respectively p=0.16 and p=0.32). Mean surgical time was 71 minute (range 51–82). Three cases of coracoid ruptures were reported, 1 failure of the system occurred.

Discussion

GPS navigation system allows the surgeon to prepare the implant site as planned on Preoperative software in Reverse shoulder arthroplasty, with no statistical difference between planned orientation and intraoperative measured orientation. That means that even in the most difficult cases the surgeon is able to find a good positioning (93% seating)and to replicate it in the operative room. Only one failure of the system occurred, because too much time was passed between CT scan and surgery (9 months). Three coracoid fractures occurred in the first 10 cases: these could be addressed to a lack of confidence with the double lateralization of this prosthesis which increase tensioning on the coracoid and a lack of confidence in tracker positioning, which should be made as proximal as it is possible. Finally, the system needs several improvements to be considered a breakthrough technology, such as humeral component positioning and final control of the implant, but by now is a useful way to improve our surgery, especially in difficult cases.