CERAMIC-ON-CERAMIC TOTAL HIP AND SIGNIFICANT TRAUMA: WHAT RISKS?

International Hip Society (IHS) Closed Meeting, Berlin, Germany, 3–5 November 2021.



Abstract

Fracture risks are the most common argument against the use of Ceramic on ceramic (CoC) hip implants. Question: is ceramic material at risk in case of severe local trauma?

Over a long period, we tried to identify patients with a CoC prosthesis (Ceraver OstealĀ°)who did sustain a trauma. This was conducted in three different institutions.

Eleven patients were found: 9 males and 2 females aged 17 to 70 years at time of index surgery. Accident occurred 6 months to 15 years after index: one car accident, five motorcycle accident, five significant trauma after a fall, including one ski board accident. Consequences of these trauma were: six fractures of the acetabulum with socket loosening in 4 that needed revision, two femoral shaft fracture, one orifed and one stem exchanged, one traumatic hip dislocation associated to loosening of the socket revised at 10 years, and one traumatic loosening of the socket. Ten had no consequence on ceramic integrity. One experienced a fracture of the patella from a dashboard trauma, a liner shipping was discovered during socket revision 2 years later. This is the only case of possible relation between trauma and ceramic fracture. In a more recent longitudinal study on 1856 CoC prosthesis performed from 2010 to 2021, 29 severe traumas were identified with no consequence on Ceramic material.

From this limited case study, it can be assumed that Pure Alumina Ceramic well designed and manufactured, will not break after a significant trauma.