Research

CERAMIC-ON-CERAMIC BEARING OUTCOMES IN TOTAL HIP ARTHROPLASTY AT A MINIMUM OF TEN YEARS

The 29th Annual Meeting of the European Orthopaedic Research Society (EORS), Rome, Italy, 15–17 September 2021.



Abstract

Introduction and Objective

Ceramic on Ceramic bearings in Total Hip Arthroplasty (THA) afford a low friction coefficient, low wear rates and extreme hardness. Significant complications include hip squeak, ceramic fracture and poor polyethylene performance in revision procedures due to imbedding of abrasive microscopic ceramic fragments. We report on the results of this bearing at a minimum of 10 years.

Materials and Methods

A single-centre retrospective review of 449 THAs was performed. Primary outcome measures included aseptic revision and all-cause revision rates at a minimum of 10 years post operatively. Evaluation of functionality was performed with WOMAC and SF-36 scores which were performed pre-operatively and at intervals of 6 months, one year, 2 years, 5 years and 10 years post operatively.

Results

There was a 6.2% (n=28) all-cause and 5.3% (n=24) aseptic revision rate for ceramic on ceramic total hip arthroplasty at minimum of 10 years with a mean time to revision 4.8 years (range 2 months − 11.6 years). Notably, there were 2 revisions for ceramic head fracture, one for ceramic liner fracture, 3 for aseptic loosening and 3 revisions for squeaking. Pain of unknown origin was the most common reason for revision. There was an improvement in postoperative WOMAC scores from a mean of 59.8 (range 15–95) pre-operatively to a mean of 15.6 (range 0–78) at 10 years.

Conclusions

This study showed good functional outcomes but high revision rates for CoC THA at a minimum of 10 years. The role for CoC bearings in THA has been called into question in recent years and may continue to decline in popularity, even in younger patients. Further large scale studies are important to assess the long-term outcomes of this bearing surface.