SIMPLE SMART IMPLANTS FOR DETECTING LOOSENING AND INFECTION

International Society for Technology in Arthroplasty (ISTA) meeting, New Early-Career Webinar Series (NEWS), held online, November 2020.



Abstract

Introduction & Aims

In other medical fields, smart implantable devices are enabling decentralised monitoring of patients and early detection of disease. Despite research-focused smart orthopaedic implants dating back to the 1980s, such implants have not been adopted into regular clinical practice. The hardware footprint and commercial cost of components for sensing, powering, processing, and communicating are too large for mass-market use. However, a low-cost, minimal-modification solution that could detect loosening and infection would have considerable benefits for both patients and healthcare providers. This proof-of-concept study aimed to determine if loosening/infection data could be monitored with only two components inside an implant: a single-element sensor and simple communication element.

Methods

The sensor and coil were embedded onto a representative cemented total knee replacement. The implant was then cemented onto synthetic bone using polymethylmethacrylate (PMMA). Wireless measurements for loosening and infection were then made across different thicknesses of porcine tissue to characterise the sensor's accuracy for a range of implantation depths. Loosening was simulated by taking measurements before and after compromising the implant-cement interface, with fluid influx simulated with phosphate-buffered saline solution. Elevated temperature was used as a proxy for infection, with the sensor calibrated wirelessly through 5 mm of porcine tissue across a temperature range of 26–40°C.

Results

Measurements for loosening and infection could be acquired simultaneously with a duration of 4 s per measurement. For loosening, the debonded implant-cement interface was detectable up to 10 mm with 95% confidence. For temperature, the sensor was calibrated with a root mean square error of 0.19°C at 5 mm implantation depth and prediction intervals of ±0.38°C for new measurements with 95% confidence.

Conclusions

This study has demonstrated that with only two onboard electrical components, it is possible to wirelessly measure cement debonding and elevated temperature on a smart implant. With further development, this minimal hardware/cost approach could enable mass-market smart arthroplasty implants.