header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

PRIMARY ROBOTIC-ARM ASSISTED TOTAL HIP ARTHROPLASTY: AN ANALYSIS OF 501 HIPS WITH 44-MONTH FOLLOW-UP

International Society for Technology in Arthroplasty (ISTA) meeting, New Early-Career Webinar Series (NEWS), held online, November 2020.



Abstract

Introduction

Primary robotic-arm assisted total hip arthroplasty (THA) yields more accurate and reproducible acetabular cup placement, nonetheless, data is scarce in terms of outcomes. The purpose of the present study was to report on patient-reported outcomes (PROMs) in a large group of patients who underwent robotic-arm assisted THA. The authors hypothesized that (1) patients who underwent robotic-arm assisted primary THA would achieve favorable and significant improvement in PROMs, (2) an accurate and reproducible acetabular cup placement with respect to the defined SafeZones would be obtained, and (3) a low rate of THA dislocation would be observed.

Methods

Prospectively collected data were retrospectively reviewed between April 2012 to May 2017. Primary THA using Mako Robotic-Arm [Mako Surgical Corp. (Stryker), Fort Lauderdale, FL, USA] with minimum two-year follow-up for the Harris Hip Score (HHS) and the Forgotten Joint Score-12 (FJS-12) were included. Exclusion criteria were: bodymass index (BMI) > 40 kg/m2, age < 21-year old, worker's compensation, or unwilling to participate. Visual analog scale (VAS) for pain and patient satisfaction were obtained. Intraoperative measurements for leg-length, global offset, acetabular inclination and version were documented.

Results

501 hips were included (57.29% females), follow-up was 43.99 ± 15.59 months. Average age was 58.70 ± 9.41 years, and the BMI was 28.41 ± 4.55 kg/m2. The group reported HHS of 90.87 ± 13.45, FJS-12 of 79.97 ± 25.87, VAS of 1.20 ± 2.06, and patient satisfaction of 8.85 ± 2.08. Intraoperative values for acetabular inclination and version were 40.0° ± 2.2 ° and 20.5° ± 2.4° respectively. Revision due to instability was 0.2%.

Conclusions

Patients who received primary robotic-arm assisted THA reported excellent results at 44-month follow-up for multiple PROMs. Consistency in acetabular cup placement accuracy was achieved in regard to the Lewinnek and Callanan safe-zones.