METAPHYSEAL CONES IN REVISION TOTAL KNEE ARTHROPLASTY WITH CEMENTED AND CEMENTLESS STEMS: EXCELLENT SURVIVORSHIP WITH SOME CAVEATS ABOUT STEM FIXATION

The Knee Society (TKS) 2020 Members Meeting, held online, 10–12 September 2020.



Abstract

Introduction

Porous metaphyseal cones are increasingly used for fixation in revision total knee arthroplasty (RTKA). Both cemented shorter length stems and longer diaphyseal engaging stems are currently utilized with metaphyseal cones with no clear evidence of superiority. The purpose of this study was to evaluate our experience with 3D printed titanium metaphyseal cones with both short cemented and longer cementless stems from a clinical and radiographic perspective.

Methods

In total 136 3D printed titanium metaphyseal cones were implanted. The mean patient age was 63 and 48% were female. The mean BMI was 33 and the mean ASA class was 2.5. There were 42 femoral cones in which 28 cemented and 14 cementless stems were utilized. There were 94 tibial cones in which 67 cemented and 27 cementless stems were utilized. The choice for stem fixation was surgeon dependent and in general cones were utilized for AORI type 2 and 3 bone defects on the femur and tibia. The most common fixation scenario was short cemented stems on both the femur and tibia followed by cemented stem fixation on the tibia and cementless fixation on the femur. Clinical data such as revision, complication, and PRO was collected at last follow-up (minimum follow-up 1 year). Radiographic analysis included cone bony ingrowth and coronal and sagittal alignment on long-standing radiographs. Descriptive statistics were used to compare demographics between patients who had malalignment (HKA beyond +/− 3 degrees and flexion/extension beyond +/− 3 degrees). Adjusted logistic regression models were run to assess malalignment risk by stem type.

Results

Patient reported outcomes demonstrated modest improvements with Pre-op KOOS improving from 44 pre-op to 59 post -op and PF-CAT improving from 33 to 37 post-op. PROMIS pain scores decreased significantly from 54 to 44 post-op.

36% of patients had malalignment in either the coronal or sagittal plane. Patients with malalignment were more likely to be female (66.7% vs 40.4%, p-value=0.02). After adjusting for age, sex and BMI, there was a significantly increased risk for coronal plane malalignment when both the femur and tibia had cementless compared to cemented stems (odds ratio=5.54, 95%CI=1.15, 26.80). There was no significantly increased risk when comparing patients with mixed stems to patients with cemented stems. Sagittal plane malalignment was more common with short cemented stems although both coronal plane and sagittal plane malalignment with either stem type was not associated with inferior clinical outcome. Overall cone survivorship was excellent with only two cones removed for infection.

Conclusion

Metaphyseal titanium cones provide reliable fixation in revision TKA. However, PROs in this complex patient population show only modest improvement consistent with other variables such as co-morbidities and poor baseline physical function. Small cone inner diameter may adversely influence cementless stem position leading to coronal plane malalignment. Short cemented stems are subject to greater sagittal plane malalignment with no apparent influence on clinical outcome.