Full Access

General Orthopaedics

TOTAL KNEE ARTHROPLASTY RADIOSTEREOMETRIC ANALYSIS IN A STANDING LOW-DOSE BIPLANAR X-RAY IMAGER

The Canadian Orthopaedic Association (COA) and The International Combined Orthopaedic Research Societies (ICORS) Meeting, Montreal, Canada, June 2019. Part 2.



Abstract

Radiostereometric analysis (RSA) allows for precise measurement of interbody distances on X-ray images, such as movement between a joint replacement implant and the bone. The low radiation biplanar EOS imager (EOS imaging, France) scans patients in a weight-bearing position, provides calibrated three-dimensional information on bony anatomy, and could limit the radiation during serial RSA studies. Following the ISO-16087 standard, 15 double exams were conducted to determine the RSA precision of total knee arthroplasty (TKA) patients in the EOS imager, compared to the standard instantaneous, cone-beam, uniplanar digital X-ray set-up.

At a mean of 5 years post-surgery, 15 TKA participants (mean 67 years, 12 female, 3 male) were imaged twice in the biplanar imager. To reduce motion during the scan, a support for the foot was added and the scan speed was increased. The voltage was also increased compared to standard settings for better marker visibility over the implant. A small calibration object was included to remove any remaining sway in post-processing.

The 95% confidence interval precision was 0.11, 0.04, and 0.15 mm in the x, y, and z planes, respectively and 0.15, 0.20, and 0.14° in Rx, Ry, and Rz. Two participants had motion artifacts successfully removed during post-processing using the small calibration object.

With faster speeds and stabilization support, this study found an in vivo RSA precision of ≤ 0.15 mm and ≤ 0.20° for TKA exams, which is within published uniplanar values for arthroplasty RSA. The biplanar imager also adds the benefits of weight bearing imaging, 3D alignment measurements, a lower radiation dose, and does not require a reference object due to known system geometry and automatic image registration.