Abstract
Mechanical alignment (MA) techniques for total knee arthroplasty (TKA) introduces significant anatomic modifications and secondary ligament imbalances. A restricted kinematic alignment (rKA) protocol was proposed to minimise these issues and improve TKA clinical results.
A total of 1000 knee CT-Scans were analyzed from a database of patients undergoing TKA. rKA tibial and femoral bone resections were simulated. rKA is defined by the following criteria: Independent tibial and femoral cuts within ± 5° of the bone neutral mechanical axis and, a resulting HKA within ± 3° of neutral. Medial-lateral (ΔML) and flexion-extension (ΔFE) gap differences were calculated and compared with MA results. With the MA technique, femoral rotation was aligned with either the trans-epicondylar axis (TEA) or with 3° of external rotation to the posterior condyles (PC).
Extension space ML imbalances (>/=3mm) occurred in 33% of TKA with MA technique versus 8% of the knees with rKA (p /=5mm) were present in up to 11% of MA knees versus 1% rKA (p < 0 .001). Using the MA technique, for the flexion space ΔML, higher imbalance rates were created by the TEA technique (p < 0 .001). rKA again performed better than both MA techniques using TEA of 3 degrees PC techniques (p < 0 .001). When all the differences between ΔML and ΔFE are considered together: using TEA there were 40.8% of the knees with < 3 mm imbalances throughout, using PC this was 55.3% and using rKA it was 91.5% of the knees (p < 0 .001).
Significantly less anatomic modifications with related ML or FE gap imbalances are created using rKA versus MA for TKA. Using rKA may help the surgeon to balance a TKA, whilst keeping the alignment within a safe range.