LEVER-OUT MOMENTS OF A LIPPED CERAMIC LINER

International Society for Technology in Arthroplasty (ISTA) meeting, 32nd Annual Congress, Toronto, Canada, October 2019. Part 2 of 2.



Abstract

Introduction

Lipped liners have the potential to decrease the rate of revision for instability after total hip replacement since they increase the jumping distance in the direction of the lip. However, the elevated lip also may reduce the Range of Motion and may lead to early impingement of the femoral stem on the liner. It is unclear whether the use of a lipped liner has an impact on the level of lever-out moments or the contact stresses. Therefore, the aim of the current study was to calculate these values for lipped liners and compare these results to a conventional liner geometry.

Materials and Methods

3D Finite Element studies were conducted comparing a ceramic lipped liner prototype and a ceramic conventional liner both made from BIOLOX®delta. The bearing diameter was 36 mm. To apply loading, a test taper made of titanium alloy was bonded to a femoral head, also made from BIOLOX®delta. Titanium was modeled with a bilinear isotropic hardening law. For the bearing contact a coefficient of friction of both 0.09 or 0.3 was assumed to model a well and poorly lubricated system. Frictionless contact was modeled between taper and liner. Pre-load was varied between 500 N and 1500 N and applied along the taper axis. While keeping pre-load constant, lever-out force was applied perpendicular to the taper axis until subluxation occurred. Liners were fixed at the taper region. Lever-out moment, equivalent plastic strain and von Mises stress of the taper, bearing contact area and contact area between taper and liner was evaluated.

Results

With increasing pre-load, larger lever-out moment, equivalent plastic strain, contact area between taper and liner and bearing contact area was found for both liner designs. However, von Mises stresses were nearly constant but slightly exceeded yield strength of titanium. For all evaluated parameters almost no differences were found between the liner designs. Lever-out moments were comparable for both designs ranging from 4.5–10.5 Nm for the lipped liner and 4.4–10.2 Nm for the conventional liner. The increase of the coefficient of friction strongly affected lever-out moments, equivalent plastic strain and contact area between taper and liner. The other parameters were not affected by varying the coefficient of friction.

Discussion

This study did not find significant differences in the lever-out behavior of the lipped acetabular liner compared to the conventional liner design. The inner geometry of the lipped liner is comparable to the conventional liner inner geometry. Therefore, contact area showed no significant differences and contact mechanics are identical in the current setup leading to similar results of both liner designs. For both designs small plastic deformations in the contact point of the taper were found at the contact region between liner and taper. However, the investigated mechanical parameters did not differ between the two investigated liner types.