ROBOTIC-ASSISTED TOTAL KNEE ARTHROPLASTY: ADDRESSING VARUS KNEES

International Society for Technology in Arthroplasty (ISTA) meeting, 32nd Annual Congress, Toronto, Canada, October 2019. Part 2 of 2.



Abstract

Introduction

Studies have shown that dissatisfaction following TKA may stem from poor component placement and iatrogenic factors related to variability in surgical execution. A CT-based robotic assisted system (RA) allows surgeons to dynamically balance the joint prior to bone resection. This study aimed to determine if this system could improve TKA planning, reduce soft tissue releases, minimize bone resection, and accurately predict component size in varus knee.

Method

Four hundred and seventy four cases with varus deformity undergoing primary RATKA were enrolled in this prospective, single center and surgeon study. Patient demographics and intraoperative surgical details were collected. Initial and final 3-dimensional alignment, component position, bone resection depths, use of soft tissue releases, knee balancing gaps, and component size were collected intraoperatively. WOMAC and KOOS Jr. scores were collected 6 months, and 1 year postoperatively. Descriptive statistics were applied to determine the changes in these parameters between initial and final values.

Results

Native deformity ranged from 1 to 19 degrees of varus. 86% of patients in this study did not require a soft tissue release regardless of their level of coronal or sagittal deformity. Complex deformities who required a soft tissue release were corrected on average to 3 degrees varus while cases without releases were corrected to 2 degrees varus on average with the overall goal as traditional mechanical alignment. All surgeons achieved their planned sizes on the tibia and femur more than 98% of the time within one size, and 100% of the time within two sizes. Flexion and extension gaps during knee balancing were within 2mm (mean 1mm) for all knees. At latest follow-up, radiographic evidence suggested well-seated and well-fixed components. Radiographs also indicated the patella components were tracking well within the trochlear groove. No revision and re-operation were reported. Mean WOMAC total score was improved from 23.8±8.0 pre-op to 8.9±7.9 1-year post-op (p<0.01). Mean KOOS Jr. score was improved from 46.8±11.6 pre-op to 77.9±14.8 1-year post-op (p<0.01).

Discussion and Conclusions

New tools may allow for enhanced execution and predictable balance for TKA, which may improve patient outcomes. In this study, preoperative planning via CT scan allowed surgeons to assess bony deformities and subtly adjust component position to reduce soft tissue trauma. While this study has several limitations, RATKA for varus knees should continue to be investigated.

For any figures or tables, please contact authors directly.