WOODPECKER PNEUMATIC FEMORAL BROACHING IN DIRECT ANTERIOR APPROACH HIP ARTHROPLASTY: HOW SAFE IS IT? AN EVALUATION OF 649 CASES

International Society for Technology in Arthroplasty (ISTA) meeting, 32nd Annual Congress, Toronto, Canada, October 2019. Part 1 of 2.



Abstract

INTRODUCTION

The Woodpecker pneumatic broaching system facilitates femoral preparation to achieve optimal primary fixation of the stem in direct anterior hip replacement using a standard operating table. The high-frequency axial impulses of the device reduce excess bone tension, intraoperative femoral fractures and overall operating time. The Woodpecker device provides uniformity and enhanced control while broaching, optimizing cortical contact between the femur and implant and thereby maximizing prosthetic axial stability and longevity. This study aims to describe a single surgeon's experience using the Woodpecker pneumatic broaching system in 649 cases of direct anterior approach (DAA) total hip arthroplasties to determine the device's safety and efficacy.

METHODOLOGY

All consecutive patients undergoing elective anterior bikini total hip arthroplasties (THA) performed by a single surgeon between July 2013 and June 2018 were included. Patients undergoing a THA with the use of the Woodpecker device through a different surgical approach, revision THA or arthroplasties for a fractured neck of femur were excluded (n=219). The pneumatic device was used for broaching the femoral canal in all cases. Pre-operative and post-operative Harris Hip Scores (HHS) and post-operative radiographs were analyzed to identify femoral fractures and femoral component positioning at 6 weeks, 6 months and 12 months post-operative. Any intra-operative or post-operative surgical complications and component survivorship until most recent follow up were recorded in the clinical notes.

RESULTS

A total of 649 patients (L THA=317, R THA=328 and bilateral=2) with a mean age of 69 (range 46–91yrs) and mean BMI of 28.3 (range = 18.4–44.0) underwent a DAA THA using a Woodpecker device were included in the study. Of these patients, 521 (80%) underwent uncemented and 128 (20%) underwent cemented femoral components. The time taken to broach the femur using Woodpecker broaching this system averaged 2.8 minutes (1.4 to 7.5 minutes) in both cemented and uncemented cases. In 91% of cases the templated broach size was achieved with the remaining 9% within +/− 1 size of the planned template.

Radiographic analysis revealed 67.3% of the stems placed in 0–1.82 degrees of varus and 32.7% placed in 0–1.4 degrees of valgus. Average HHS were 24.4 pre-operatively, with drastic improvements shown at 6 weeks (80.95), 6 months (91.91) and 12 months (94.18) after surgery. Intraoperative femoral fractures occurred in three patients (0.4%) during trial reduction, a further three patients had periprosthetic post-operative fractures (0.4%) from falls, two patients had stem subsidence (0.3%) and a further two patients had wound infections (0.3%). At the most recent follow up, the survivorship of the acetabular component was 99.7% and the femoral component was 99.1%, with mean follow up of 2.9 years (0.5 to 5 years). No intraoperative or post-operative complications could be directly attributed to the Woodpecker broaching system.

CONCLUSION

The pneumatic Woodpecker device is a safe and effective alternative tool in minimally invasive direct anterior hip replacement surgery for femoral broaching performed on a standard operating table. The skill and experience of the surgeon must be taken into consideration when utilizing new surgical devices.