header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

KINEMATIC VALIDATION OF A SURFACE-GUIDED ARTICULATION CONCEPT FOR TOTAL KNEE ARTHROPLASTY

International Society for Technology in Arthroplasty (ISTA) meeting, 32nd Annual Congress, Toronto, Canada, October 2019. Part 1 of 2.



Abstract

Background

The anatomy of the human knee is very different than the tibiofemoral surface geometry of most modern total knee replacements (TKRs). Many TKRs are designed with simplified articulating surfaces that are mediolaterally symmetrical, resulting in non-natural patterns of motion of the knee joint [1]. Recent orthopaedic trends portray a shift away from basic tibiofemoral geometry towards designs which better replicate natural knee kinematics by adding constraint to the medial condyle and decreasing constraint on the lateral condyle [2]. A recent design concept has paired this theory with the concept of guided kinematic motion throughout the flexion range [3]. The purpose of this study was to validate the kinematic pattern of motion of the surface-guided knee concept through in vitro, mechanical testing.

Methods

Prototypes of the surface-guided knee implant were manufactured using cobalt chromium alloy (femoral component) and ultra-high molecular weight polyethylene (tibial component). The prototypes were installed in a force-controlled knee wear simulator (AMTI, Watertown, MA) to assess kinematic behavior of the tibiofemoral articulation (Figure 1). Axial joint load and knee flexion experienced during lunging and squatting exercises were extracted from literature and used as the primary inputs for the test. Anteroposterior and internal-external rotation of the implant components were left unconstrained so as to be passively driven by the tibiofemoral surface geometry. One hundred cycles of each exercise were performed on the simulator at 0.33 Hz using diluted bovine calf serum as the articular surface lubricant. Component motion and reaction force outputs were collected from the knee simulator and compared against the kinematic targets of the design in order to validate the surface-guided knee concept.

Results

Under deep flexion conditions of up to 140° of squatting the surface-guided knee implants were found to undergo a maximum of 22.2° of tibial internal rotation and 20.4 mm of posterior rollback on the lateral condyle. Pivoting of the knee joint was centered about the highly congruent medial condyle which experienced only 1.6 mm of posterior rollback. Experimental results were within 2° (internal-external rotation) and 1 mm (anteroposterior translation) agreement with the design target throughout the applied exercises (Figure 2).

Conclusion

The results of this test confirm that by combining a constrained medial condyle with guiding geometry on the lateral condyle, deep knee flexion activities of up to 140° can be performed while maintaining near-natural kinematics of the knee joint. The authors believe that the tested surface-guided implant concept is a significant step toward the development of novel TKR which allows a greater range of motion and could improve the quality of life for active patients undergoing knee replacement.

For any figures or tables, please contact the authors directly.