Introduction

Biological repair of acetabular bone defects after impaction bone grafting (IBG) in total hip arthroplasty could facilitate future re-revisions in case of failure of the reconstruction again using the same technique. Few studies have analysed the outcome of these acetabular re-revisions.

Patients and Methods

We analysed 34 consecutive acetabular re-revisions that repeated IBG and a cemented cup in a cohort of 330 acetabular IBG revisions. Fresh-frozen femoral head allografts were morselized manually. All data were prospectively collected. Kaplan-Meier survivorship analysis was performed. The mean follow-up after re-revision was 7.2 years (2–17). Intraoperative bone defect had lessened after the first failed revision. At the first revision there were 14 hips with Paprosky 3A and 20 with Paprosky type 3B. At the re-revision there were 5 hips with Paproky 2B, 21 with Paprosky type 3A and 8 with type 3B. Lateral mesh was used in 19 hips.

Results

The mean Harris Hip Score improved from 45.4 (6.7) to 77.1 (15.6) at final follow-up. The radiological analysis showed cup migration in 11 hips. The mean appearance time was 25 months (3–72). Of these, migration in three cups was progressive and painful requiring re-revision. Cup tilt was found in all migrated hips. There were one dislocation requiring a cemented dual mobility cup associated with IBG and one infection resolved with resection-arthroplasty. Survival with further cup revision for aseptic loosening was 80.7% (95% Confidence Interval 57.4–100) at 11 years. In all surviving re-revisions trabecular incorporation was observed without radiolucent lines.

Conclusion

Biological repair can be obtained by restoring the bone stock, even after successive acetabular reconstructions using IBG and a cemented cup.