REMOVING A WELL-FIXED ACETABULAR COMPONENT: BLADE RUNNER 2018

The Current Concepts in Joint Replacement (CCJR) Spring 2018 Meeting, Las Vegas, NV, USA, 20–23 May 2018.



Abstract

A well-fixed uncemented acetabular component is most commonly removed for chronic infection, malposition with recurrent dislocation, and osteolysis. However, other cups may have to be removed for a broken locking mechanism, a bad “track record”, and for metal-on-metal articulation problems. Modern uncemented acetabular components are hemispheres which have 3-dimensional ingrowth patterns. Coatings include titanium or cobalt-chromium alloy beads, mesh, and now the so-called “enhanced coatings”, such as tantalum trabecular metal, various highly porous titanium metals, and 3-D printed metal coatings. These usually pose a problem for safe removal without fracture of the pelvis or creation of notable bone deficiency.

Preoperative planning is essential for safe and efficient removal of these well-fixed components. Strongly consider getting the operative report, component “stickers”, and contacting the implant manufacturer for information. There should a preoperative check list of the equipment and trial implants needed, including various screwdrivers, trial liners, and a chisel system. The first step in component removal is excellent 360-degree exposure of the acetabular rim, and this can be accomplished by several approaches. Then, the acetabular polyethylene liner is removed; a liner that is cemented into a porous shell can be “reamed out” using a specific device. Following this, any central or peripheral screws are removed; broken or stripped screw heads add an additional challenge. A trial acetabular liner is placed, and an acetabular curved chisel system is used. There are two manufacturers of this type of system. Both require the known outer acetabular diameter and the inner diameter of the trial liner. With the curved chisel system and patience, well-fixed components can be safely removed, and the size of the next acetabular component to be implanted is usually 4mm larger than the one removed. There are special inserts for removal of monobloc metal shells. Remember that removal of these well-fixed components is more difficult in patients compared to models, and is just the first step of a successful acetabular revision.