Abstract

Total knee replacements are being more commonly performed in active younger and obese patients. Fifteen-year survivorship studies demonstrate that cemented total knee replacements have excellent survivorship, with reports of 85 to 97%. Cemented knee arthroplasties are doomed to failure due to loss of cement-bone interlock over time. Inferior survivorship occurs in younger patients and obese patients who would be expected to place increased stress on the bone-cement interfaces. Roentgen stereophotogrammetric analysis (RSA) studies have indicated that cementless fixation should perform better than cemented fixation. However, cementless fixation for total knee replacement has not gained widespread utilization due to the plethora of poor results reported in early series. The poor initial results with cementless total knee replacement have occurred due to poor implant designs such as cobalt chrome porous interfaces, poor initial tibial component stability, lack of continuous porous coating, poor polyethylene, and use of metal-backed patellae.

I have used cementless fixation for total knee replacements for young, active, and heavy patients since 1986 when durability over 20 years is desirable. My series of over 1,300 cementless TKAs represents about 20% of the 6,500 total knees I have performed from 1986 to 2017. I have seen initial failures in my series due to the use of metal-backed patellae with thin polyethylene, older generation polyethylene, and use of screws with the tibial components which provide access to the metaphyseal bone for polyethylene wear debris. Overall implant fixation failures were still significantly low due to the use of a highly porous titanium surface on both the tibial and femoral components. With the advent of utilizing implants with continuous porous surfaces and highly crosslinked polyethylene, and elimination of use of metal-backed patellae and tibial screws, I have only had one revision due to aseptic loosening or osteolysis in the last 1,071 cases performed since 2002.

Almost 50% of total knees are now performed on patients under the age of 65. A 55-year-old patient has a 30 year life expectancy. Modern total knee replacement design has made biological fixation predictable for young and heavy patients. Because it is a biological interface, it should respond better than cement to the increased stresses that will be applied over many years by our younger, more active and heavier total knee population.