EFFECT OF SOFT-TISSUE CONSTRAINTS AND COMPONENT ALIGNMENT ON THE KINEMATICS AND WEAR OF TOTAL KNEE ARTHROPLASTIES

International Society for Technology in Arthroplasty (ISTA) 31st Annual Congress, London, England, October 2018. Part 2.



Abstract

Component alignment and soft tissue constraints are key factors affecting function and implant survival after total knee replacement (TKR). Knee kinematics contribute to knee function whilst soft tissue constraints and component alignment impact polyethylene wear. This study experimentally investigated the effect of soft tissue constraints and component alignment on the kinematics and wear of a TKR.

A six station electromechanical ProSim knee simulator was used with the ISO 14243-1:2009 standard force control inputs; axial force, flexion-extension (FE), tibial rotation (TR) torque and anterior-posterior (AP) force. This allowed the kinematics to vary with the test conditions. The soft tissue constraints were simulated using virtual springs.

DePuy Sigma XLK fixed bearing TKRs were tested in 25% bovine serum (in 0.04% sodium azide) lubricant. The average output kinematics across 6 stations were found for each test and the peak values compared. The wear rates were calculated over 2 million cycles (MC), the serum was changed every 350,000 cycles and the tibial inserts weighed after every MC. A one way ANOVA and post hoc Tukey's test was used to compare the kinematics and wear with significance taken at p<0.05.

The kinematics and wear rates for three soft tissue conditions were established under ideal alignment (Table 1). The ISO standard springs for a cruciate substituting (CS) and a cruciate retaining (CR) prosthesis were used to represent a knee with a resected ACL and PCL and a knee with a resected ACL respectively. The third spring condition was based on clinical data to represent a “stiff” knee.

Three other alignment conditions were then assessed using “stiff” knee springs; 4° varus, 14° rotational mismatch and 10° posterior tibial slope. These alignments were chosen to represent the range found in clinical data.

Under ideal alignment the “stiff” knee springs had significantly lower peak AP and TR displacements (0.9mm, 2mm, 2mm and 3.6°, 7.1°, 7.8° for the “stiff”, CR and CS springs respectively) than the other springs (p<0.01). The “stiff” knee spring had a significantly lower wear rate than the CR spring; 1.58 ±1.20mm³/MC compared to 4.71±1.29 mm³/MC (p<0.01).

The varus and rotated components had significantly larger peak AP displacements of 2.56mm and 2.42mm respectively, than the ideal and tibial slope fixtures (1.97mm and 0.92mm respectively) (p<0.01). The rotated components had significantly higher internal rotation of 12.2° compared to 4.4°, 3.7° and 3.5° for the tibial slope, varus and ideal components respectively (p<0.01).

The ideal and varus components had significantly lower wear than the tibial slope and rotated components (1.58±1.20mm³/MC and 0.15±0.83mm³/MC compared to 8.24±7.72mm³/MC and 5.19±1.12mm³/MC respectively) (p<0.01). This may be due to increased AP and TR displacements with the rotated components and the increased anterior AP displacement with the tibial slope components, resulting in wear on the posterior edge of the tibial insert.

Soft tissue constraints and component alignment had a significant effect on the kinematics and wear. Experimental simulation should test a variety of soft tissue and alignment conditions to reflect the range observed clinically and determine causes for early failure.

For any figures or tables, please contact the authors directly.