header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

ARE SUPINE INTRAOPERATIVE AND STANDING BIPLANAR POSTOPERATIVE RADIOLOGICAL MEASUREMENTS OF ACETABULAR CUP POSITION THE SAME?

International Society for Technology in Arthroplasty (ISTA) 31st Annual Congress, London, England, October 2018. Part 1.



Abstract

Background

Supine positioning during direct anterior approach total hip arthroplasty (DAA THA) facilitates use of fluoroscopy, which has been shown to improve acetabular component positioning on plane radiograph. This study aims to compare 2- dimensional intraoperative radiographic measurements of acetabular component position with RadLink to postoperative 3- dimensional SterEOS measurements.

Methods

Intraoperative fluoroscopy and RadLink (El Segundo, CA) were used to measure acetabular cup position intraoperatively in 48 patients undergoing DAA THA. Cup position was measured on 6-week postoperative standing EOS images using 3D SterEOS software and compared to RadLink findings using Student's t-test. Safe-zone outliers were identified. We evaluated for measurement difference of > +/− 5 degrees.

Results

RadLink acetabular cup abduction measurement (mean 43.0°) was not significantly different than 3D SterEOS in the anatomic plane (mean 42.6°, p = 0.50) or in the functional plane (mean 42.7°, p = 0.61) (Fig. 1–2). RadLink acetabular cup anteversion measurement (mean 17.9°) was significantly different than 3D SterEOS in both the anatomic plane (mean 20.6°, p = 0.022) and the functional plane (mean 21.2°, p = 0.002) (Fig. 3–4). RadLink identified two cups outside of the safe-zone. However, SterEOS identified 12 (anatomic plane) and 10 (functional plane) outside of the safe-zone (Fig. 5–7). In the functional plane, 58% of anteversion and 92% of abduction RadLink measurements were within +/− 5° of 3D SterEOS.

Conclusion

Intraoperative fluoroscopic RadLink acetabular anteversion measurements are significantly different than 3D SterEOS measurements, while abduction measurements are similar. Significantly more acetabular cups are placed outside of the safe- zone when evaluated with 3D SterEOS versus RadLink.


Email: