IN VIVO MEASUREMENT OF DISTANCE BETWEEN SCAPULAR NECK AND POLYETHYLENE INSERT DURING ACTIVE EXTERNAL ROTATION IN SHOULDERS WITH GRAMMONT TYPE REVERSE PROSTHESIS

International Society for Technology in Arthroplasty (ISTA) 31st Annual Congress, London, England, October 2018. Part 1.



Abstract

Background

Scapular notching is a complication after reverse shoulder arthroplasty with a high incidence up to 100%. Its clinical relevance remains uncertain; however, some studies have reported that scapular notching is associated with an inferior clinical outcome. There have been no published articles that studied positional relationship between the scapular neck and polyethylene insert in vivo. The purpose of this study was to measure the distance between the scapular neck and polyethylene insert in shoulders with Grammont type reverse shoulder arthroplasty during active external rotation at the side.

Methods

Eighteen shoulders with Grammont type prosthesis (Aequalis Reverse, Tornier) were enrolled in this study. There were 13 males and 5 female, and the mean age at surgery was 74 years (range, 63–91). All shoulders used a glenosphere with 36mm diameter, and retroversion of the humeral implant was 10°in 4 shoulders, 15°in 3 shoulders, and 20°in 11 shoulders. Fluoroscopic images were recorded during active external rotation at the side from maximum internal to external rotation at the mean of 14 months (range, 7–24) after surgery. The patients also underwent CT scans, and three-dimensional glenosphere models with screws and scapula neck models were created from CT images. CT-derived models of the glenosphere and computer-aided design humeral implant models were matched with the silhouette of the implants in the fluoroscopic images using model-image registration techniques (Figure 1). Based on the calculated kinematics of the implants, the closest distance between the scapular neck and polyethylene insert was computed using the scapular model and computer-aided design insert models (Figure 2). The distance was computed at each 5° increment of glenohumeral internal/external rotation, and the data from 20°internal rotation to 40°external rotation were used for analyses. One-way repeated-measures analysis of variance was used to examine the change of the distance during the activity, and the level of significance was set at P < 0.05.

Results

The mean glenohumeral abduction during the activity was 17°-22°. The mean distance between the neck and insert was approximately 1mm throughout the activity (Figure 3). The distance tended to become smaller with the arm externally rotated, but the change was not significant.

Discussion

The reported incidence of scapular notching after Grammont type reverse shoulder arthroplasty is generally higher than the newer design prosthesis with the lateralized center of rotation. This may be associated with the design of the prosthesis, and the results of this study that the distance between the neck and insert was approximately 1mm throughout active external rotation at the side will support the high incidence of notching. We may need to analyze the distance with the newer design reverse shoulder prosthesis to prove the architectural advantage of the newer systems.

Conclusion

The distance between the scapular neck and polyethylene insert was approximately 1mm throughout active external rotation activity in shoulders with Grammont type prosthesis.