POSTERIOR LOADING DURING DEEP KNEE BEND: A CRITICAL BIOMECHANICAL STATE FOR EVALUATING BONE FILLERS IN REVISION TOTAL KNEE ARTHROPLASTY

International Society for Technology in Arthroplasty (ISTA) 31st Annual Congress, London, England, October 2018. Part 1.



Abstract

INTRODUCTION

Porous metal bone fillers are frequently used to manage bony defects encountered in revision total knee arthroplasty (rTKA). Compared to structural graft, porous metal bone fillers have shown significantly lower loosening and failure rates potentially due to osseointegration and increased material strength [1]. The strength of porous metal bone fillers used in lower extremities is frequently assessed using compression/shear/torsion test methods, adapted from spine standards. However, these basic methods may lack clinical relevance, and do not provide any insight on the relationship between patient activity and anticipated prosthesis performance. The goal of this study was to evaluate the response of bone fillers under different activities of daily living, in order to define physiologically relevant worst case biomechanics for component evaluation.

METHODS

A bone filler tibial augment is shown in Figure 1. A test construct for tibial augments (half-block each for medial and lateral sides) is shown in Figure 2, along with compatible rTKA components. An additional void in the bone was filled using bone cement. Loading was applied through the tibiofemoral contact patches created on polyethylene tibial insert. Loading was used for two activities of daily living; walking and deep knee bend [2–3]. During walking, the tibiofemoral contact patch on the anterior tibial post gets loaded due to femoral hyperextension with 1.2xbody weight (BW), whereas the medial and lateral condyles get loaded with 3xBW compressive load. For deep knee bend, only the condyles get loaded with 4.34xBW. Compared to walking, 45% higher compressive load magnitude in deep knee bend located further posterior was anticipated to create a larger bending moment and induce higher stress on the half augments. A finite element analysis (FEA) was performed by modeling this test construct with a medium size tibial augment. All components were modeled using linear elastic material properties. All interfaces, including the augment-bone interface (representing full bony ingrowth construct) were modeled using bonded contact. The inferior surface of the bone analogue was constrained. Linear static analyses were performed and peak von mises stress predicted in the tibial augments was compared between activities.

RESULTS

Deep knee bend resulted in 31% higher stresses in the tibial augments than for walking. High von mises stresses were mostly predicted at the superior/posterior aspect of the internal side of the augment and in the corners of the cutouts. Figure 3 presents the von mises stresses in the tibial augments for both loading scenarios.

DISCUSSION

This study revealed that the 45% increased posterior compressive load associated with deep knee bend is a more significant factor than the moment applied to the post during walking gait for a hyperextended knee, when considering the stress in bone filler augments in revision TKA. The stress in the augments can depend on multiple factors and the proposed FEA method can be used to compare stresses in different porous material bone fillers to determine worst case for assessing its strength.