CEMENTING LINERS INTO WELL-FIXED ACETABULAR COMPONENTS: SHOULD WE BE CONCERNED ABOUT LONG-TERM FIXATION?

The Hip Society (THS) 2019 Summer Meeting, Kohler, WI, USA, 25–27 September 2019.



Abstract

Introduction

Cementation of a new liner into an existing well-fixed acetabular component is common during revision total hip arthroplasties (THAs) for many indications, but most commonly for lack of a modern compatible crosslinked polyethylene liner. However, little is known about the long-term durability of this strategy. The purpose of this study was to evaluate the long-term implant survivorship, risk of complications, clinical outcomes, and radiographic results of cementing a new highly cross-linked polyethylene (HXLPE) liner into a well-fixed acetabular component.

Methods

We retrospectively identified 326 revision THAs where a non-constrained HXLPE liner was cemented into a well-fixed acetabular component. Mean age at revision THA was 63 years, with 50% being female. The most common indications for revision THA were wear and osteolysis (49%), aseptic femoral loosening (35%), and instability (8%). Mean follow-up was 10 years.

Results

Polyethylene liner failure occurred in 15 cases (5%). In all cases, the cemented liner dissociated from the acetabular component. Survivorships free from any revision and any reoperation were 79% and 77% at 10 years, respectively. The most common reason for re-revision was dislocation (56% of re-revisions). The cumulative incidence of dislocation was 17% at 10 years. Hips revised at the index revision for instability were significantly more likely to have a subsequent dislocation when compared to those revised for polyethylene liner wear (HR 2.5, p<0.01). Harris hip scores significantly improved from a mean of 65 preoperatively to 88 postoperatively (p < 0.01).

Conclusions

Cementation of a non-constrained HXLPE liner into a well-fixed acetabular component during revision THA provided durable fixation at 10 years with only a small number of failures at the cement interface (5%). Instability after this procedure remains a concern, but this is likely multi-factorial in nature. These new long-term data support continued use of this technique, when necessary, during revision THAs.