header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

General Orthopaedics

EXPERIMENTAL CHARACTERIZATION OF THE ACETABULAR CUP FIXATION STABILITY IN DEPENDENCE ON INCREASING ACETABULAR DEFECT SITUATION

The International Society for Technology in Arthroplasty (ISTA), 30th Annual Congress, Seoul, South Korea, September 2017. Part 2 of 2.



Abstract

Introduction

Modern acetabular cups require a convenient bone stock for sufficient cup fixation. Thereby, fixation stability is influenced by the chosen interference fit of the acetabular cup, the cup surface structure, circularity of the reamed acetabulum and by the acetabular bone quality. The ideal implantation situation of the cup is commonly compromised by joint dysplasia and acetabular bone defects. The aim of the present experimental study was to characterise implant fixation of primary acetabular cups in case of definite acetabular cavity defects.

Materials and Methods

For the experimental determination bone substitute blocks (100 × 100 × 50 mm) made of polymethacrylimide (PMI) foam with a density of 7 pcf were used. The created acetabular defect situations were derived from the defect classification according to Paprosky. The defect geometries in the PMI foam blocks were realised by a CNC drilling machine. Thereby the defects are described in the dorso-ventral direction by the angle α and in medio-lateral direction by the angle β (given as angle combination α/β) related to the centre of rotation of the reamed cavity. For the lever-out tests the defect types IIb and IIIa (each with different α and β angles) were considered and compared to the intact fixation situation. Therefore, a macrostructured titanium cup (Allofit, Zimmer GmbH, Wintherthur, Switzerland) with an outer diameter of 56 mm were displacement-controlled (v = 20 mm/min) pushed into the 2 mm diametric under reamed PMI-foam cavities. Three cups were inserted until the cup overhang pursuant to surgical technique was reached. Subsequently the cups were displacement-controlled (v = 20 mm/min) levered out via a rod which was screwed into the implant pole by perpendicular displacement (Uaxial) of the rod in direction of the defect aperture. The lever-out moments were calculated by multiplying the first occurring force maximum (Fmax) with the effective lever arm length (llever), whereby moments caused by the deadweight of the rod were considered. Primary stability was defined by the first maximum lever-out moment.

Results

The calculated lever-out moments were in a range from 15.5 ± 1.4 Nm to 1.4 ± 0.5 Nm. Defects with a 90° dorso-ventral opening angle showed 57 ± 17% lower lever-out moments. Defects with a 120° dorso-ventral opening angle showed 80 ± 6% lower lever-out moments compared to the cup fixation into intact cavities. Moreover, medio-lateral angles greater than 20° reduced the lever-out moment by 79 ± 12% compared to the intact cavities.

Conclusion

The determined lever-out moments underline the reduction of fixation stability of acetabular cup by loss of circumferential rim and absent of superior wall support of the acetabular bone. Thereby, the fixation stability is influenced by the degree of dorso-ventral and medio-lateral defect manifestation. Hence, the fixation stability depends on the cavity surface and in particular the surface of the bone-implant interface in the fixation zone of the acetabular cup Thus, dorso-ventral defect sizes with greater opening angle than 60° and medio-lateral defect sizes greater than 20° are critically for sufficient fixation of primary acetabular cup implants.


Email: