POLYETHYLENE INSERT SUBLUXATION IN ROTATING PLATFORM TOTAL KNEE ARTHROPLASTY: THE ROLE OF FLEXION/EXTENSION GAP LAXITY AND ASYMMETRY

The Knee Society (TKS) 2018 Members Meeting, Saint Louis, MO, USA, September 2018.



Abstract

Introduction

Clinical observations suggest mid-flexion instability may occur more commonly with rotating platform (RP) total knee arthroplasty (TKA), including increased revision rates and patient-reported instability and pain. We propose that increased gap laxity leads to liftoff of the lateral femoral condyle with decreased conformity between the femoral component and polyethylene (PE) insert surface leading to PE subluxation or dislocation. The objectives of this study were to define “at risk” loading conditions that predispose patients to PE insert subluxation or spinout, and to quantify the margin of error for flexion/extension gap laxity in preventing these adverse events under physiologic loading conditions.

Methods

Biomechanical testing was performed on six fresh frozen cadaveric knees implanted with a posterior stabilized RP TKA using a gap balancing technique. Rotational displacement and torque were measured over time, while stiffness, yield torque, max torque and displacement were calculated using a post-processing, custom MatLab code. Revision with varying size femoral components (size 3–6) and PE insert thicknesses (10–15mm), by downsizing one step, were used to create a spectrum of flexion/extension gap mismatch. Each configuration was subjected to three loaded testing conditions (0°, 30° and 60° flexion) in balanced and eccentric varus loading, known to represent daily clinical function and “at risk” circumstances.

Results

PE insert rotational instability was primarily determined by conformity and contact area between the femoral condyle and the upper surface of the PE insert. In this RP design, contact area is known to decrease with flexion greater than 35°, which predisposed to abnormal motion of the femur on PE insert (Figure 1). Under all flexion/extension gap testing conditions, PE insert rotational displacement significantly decreased with increasing knee flexion (differences ranged from 0.42 to 1.01cm, p<0.05), confirming that decreased conformity allows unintended motion to occur on the upper rather than the lower insert surface, as kinematically designed. This decrease in insert rotation was further exacerbated with eccentric medial-sided loading (differences ranged from 0.77 to 1.18cm, p<0.05). Yield torque (19.66±6.79N-m, p=0.033) and max torque (19.76±5.93N-m, p=0.014) significantly increased with increasing flexion from 0° to 60° under gap balanced conditions. Yield torque significantly decreased with greater flexion gap laxity at 60° of flexion (−24.82±5.96N-m, p=0.004). The depth of the lateral PE insert concavity (1.7–3.6mm) varied with insert size and thickness and determined femoral condylar capture. The lateral insert concavity defines a narrow margin of error in flexion/extension gap asymmetry leading to rotational insert instability, especially in smaller sized knees (size 3) where the jump height (1.7mm) is less than the insert sizing increment of 2.5mm.

Conclusions

Contact area is known to decrease with flexion greater than 35° in this TKA-RP design. Flexion gap laxity further increased the risk of unintended top-side rotation of the femur on the insert, especially with increasing flexion and smaller components. In RP-TKA, in addition to medial-lateral gap symmetry and flexion-extension balance, a snug flexion gap with less than 2mm lateral laxity is critical to avoid insert instability and condylar escape with insert subluxation.

For any figures or tables, please contact authors directly.