Knee

EVALUATION OF HIGHLY POROUS BIOLOGICAL FIXATION IN TOTAL KNEE ARTHROPLASTY USING RADIOSTEREOMETRIC ANALYSIS: ASSESSMENT OF TIBIA BASEPLATE AND METAL-BACKED PATELLA

The Knee Society (TKS) 2018 Members Meeting, Saint Louis, MO, USA, September 2018.



Abstract

Introduction

Uncemented highly porous titanium implants have been shown to promote osseointegration, and may result in a durable construct for total knee arthroplasty (TKA). Given the mixed results of uncemented TKA, it is important to evaluate the early stability for this product. The objective of the following study was to use radiostereometric analysis (RSA) to assess early fixation of a highly porous tibial baseplate and metal backed patella.

Methods

Twenty-seven patients (mean age 64 years, 30% female) undergoing primary TKA consented to participate in this prospective cohort study. All patients received a highly porous tibial baseplate, a metal backed patella and tantalum RSA bone markers. Implant migration was assessed using model-based RSA at 1.5, 3, 6, 12 and 24 months post-operative. Patient reported outcome measures were captured using the same follow-up schedule, and compared to pre-operative measures.

Results

There were no adverse events affecting implant fixation, and no revisions. Patient function significantly improved by 3 months post-operation (p < 0.001). The highest rate of tibia and patellar component migration occurred over the first six post-operative weeks, with minimal migration thereafter. Mean maximum total point motion (MTPM) at 24 months was 0.72 (SD 0.34) mm for the tibia, and 0.44 (SD 0.25) mm for the patella. Three tibia baseplates migrated more than 1 standard deviation greater than the mean at 24 months, and also had continuous migration (> 0.2mm of MTPM) in the second post-operative year. One patellar component showed a rapid rate of migration between 6 and 24 months, whereas all other patellar components appeared to stabilize.

Conclusions

Osseointegration appears to occur on the highly porous implant surface of the tibia baseplate and metal backed patella, as evidenced by implant stability. Further follow-up is required to determine if clinical loosening will manifest in the continuously migrating implants.