Abstract

The goals of revision total hip on the femoral side are to achieve long term stable fixation, improve quality of life and minimise complications such as intra-operative fracture or dislocation. Ideally these stems will preserve or restore bone stock. Modular titanium stems were first introduced in North America around 2000. They gained popularity as an option for treating Paprosky 3B and 4 defects.

Several studies at our institution have compared modular titanium stems with monoblock cobalt chromium stems. The main outcomes of interest were quality of life. We also looked at complications such as intra-operative fracture and post-operative dislocation. We also compared these 2 stems with respect to restoration or preservation of bone stock. In 2 studies we showed that modular titanium stems gave superior functional outcomes as well as decreased complications compared to a match cohort of monoblock cobalt chromium stems.

As mentioned, one of the initial reasons for introduction of these stems was to address larger femoral defects where failure rates with monoblock cobalt chromium stems were unacceptably high. We followed a group of 65 patients at 5–10 years post revision with a modular fluted titanium stem. Excellent fixation was obtained with no cases of aseptic loosening. However, there were 5 cases of fracture of the modular junction.

Due to concerns of fracture of the modular junction, more recently, at our institution, we have switched to almost 100% monoblock fluted titanium stems. We recently reviewed our first 100 cases of femoral revision with a monoblock stem. Excellent fixation was achieved with no cases of aseptic loosening. Quality of life outcomes were similar to our previous reported series on modular tapered titanium stems.

Both monoblock and modular fluted titanium stems can give excellent fixation and excellent functional outcomes. This leaves a choice for the surgeon. For the low volume revision surgeon modular tapered stems are probably the right choice.

Higher volume surgeons or surgeons very comfortable with performing femoral revision may want to consider monoblock stems. If one is making the switch it would be easiest to start with a simple case. Such a case would be one that can be done with a endofemoral approach. In this approach the greater trochanter is available as the key landmark for reaming. After the surgeon is comfortable with this stem more complex cases can easily be handled with the monoblock stem.

In summary, both modular and monoblock titanium stems are excellent options for femoral revision. As one becomes more familiar with the monoblock stem it can easily become your workhorse for femoral revision. At our institution we introduced a monoblock titanium stem in 2011. It started out at 50% of cases and now it is virtually used in almost 100% of revision cases.