ROBOT-BONE ATTACHMENT DEVICE FOR ROBOT-ASSISTED PERCUTANEOUS BONE FRAGMENT MANIPULATION

International Society of Computer-Assisted Orthopaedic Surgery (CAOS), 17th Annual Scientific Meeting, Aachen, June 2017



Abstract

The treatment of joint-fractures is a common task in orthopaedic surgery causing considerable health costs and patient disabilities. Percutaneous techniques have been developed to mitigate the problems related to open surgery (e.g. soft tissue damage), although their application to joint-fractures is limited by the sub-optimal intra-operative imaging (2D- fluoroscopy) and by the high forces involved. Our earlier research toward improving percutaneous reduction of intra-articular fractures has resulted in the creation of a robotic system prototype, i.e. RAFS (Robot-Assisted Fracture Surgery) system.

We propose a robot-bone attachment device for percutaneous bone manipulation, which can be anchored to the bone fragment through one small incision, ensuring the required stability and reducing the “biological cost” of the procedure. It consists of a custom-designed orthopaedic pin, an anchoring system (AS secures the pin to the bone), and a gripping system (GS connects the pin and the robot). This configuration ensures that the force/torque applied by the robot is fully transferred to the bone fragment to achieve the desired anatomical reduction.

The device has been evaluated through the reduction of 9 distal femur fractures on human cadavers using the RAFS system. The devices allowed the reduction of 7 fractures with clinical acceptable accuracy. 2 fractures were not reduced: in one case the GS failed and was not able to keep the pin stationary inside the robot (pin rotates inside the GS). The other fracture was too dislocated (beyond the operational workspace capability of the robot). A more stable GS will be designed to avoid displacements between the pin and the robot.