OUTCOME OF HINGE KNEE ARTHROPLASTY FOR COMPLEX KNEE REVISIONS

British Indian Orthopaedic Society (BIOS) (previously IOSUK) Annual Scientific Meeting, July 2017



Abstract

Background

One of the major concerns of hinge knees have been reported in literature is mechanical failure. Failure in the form of component fracture (2–10%) and hinge dislocation/ failure are worrisome. In addition, higher risk of aseptic loosening with hinge knee prosthesis has been attributed to stress transfer at bone cement interface.

Methods

Retrospective review of clinical and radiological results of 71 consecutive patients operated at single centre using Smiles hinge knee (Stanmore implants) between 2010 and 2014. Data was collected till the latest follow up. Mechanical failure due to any reason was considered as primary end point. Radiological evidence of aseptic loosening was considered to be one of the surrogate end points.

Results

Mean age at revision knee was 65.4 years (range 42 – 84). The number of previous revisions ranged from 1 – 4. Most common indications for revision were infection (n=26) and instability (n=26). Majority of patients had rotating hinge knee (n=66). Average follow-up was 31 months (range 3 – 72). Four patients died during follow-up due to medical reasons. There were no mechanical failures of implants apart from backing out of axle due to failure of circlip in 1 patient. Asymptomatic aseptic loosening of implant on femoral side was noted in 1 patient; however, there were no peri-prosthetic fractures in this series. Three patients had re-revision for infection and one for extensor mechanism subluxation.

Conclusion

To our knowledge, this is the largest series till date reporting on mechanical outcomes of hinged knee implants for complex revisions. Clip failure is of concern; however there have been no other mechanical failures. Early results are encouraging, with no symptomatic aseptic loosening of implants to report.


Email: