The in vivo effect of a porous titanium alloy flange with hydroxyapatite, silver and fibronectin coatings on soft-tissue integration of intraosseous transcutaneous amputation prostheses



Download PDF

Abstract

Aims

The Intraosseous Transcutaneous Amputation Prosthesis (ITAP) may improve quality of life for amputees by avoiding soft-tissue complications associated with socket prostheses and by improving sensory feedback and function. It relies on the formation of a seal between the soft tissues and the implant and currently has a flange with drilled holes to promote dermal attachment. Despite this, infection remains a significant risk. This study explored alternative strategies to enhance soft-tissue integration.

Materials and Methods

The effect of ITAP pins with a fully porous titanium alloy flange with interconnected pores on soft-tissue integration was investigated. The flanges were coated with fibronectin-functionalised hydroxyapatite and silver coatings, which have been shown to have an antibacterial effect, while also promoting viable fibroblast growth in vitro. The ITAP pins were implanted along the length of ovine tibias, and histological assessment was undertaken four weeks post-operatively.

Results

The porous titanium alloy flange reduced epithelial downgrowth and increased soft-tissue integration compared with the current drilled flange. The addition of coatings did not enhance these effects.

Conclusion

These results indicate that a fully porous titanium alloy flange has the potential to increase the soft-tissue seal around ITAP and reduce susceptibility to infection compared with the current design.

Cite this article: Bone Joint J 2017;99-B:393–400.

The Intraosseous Amputation Prosthesis (ITAP) (Stanmore Implants Worldwide Ltd, Elstree, United Kingdom) is an osseointegrated, transcutaneous titanium alloy implant that penetrates the skin to attach an external prosthesis. It avoids many of the common problems of conventional socket prostheses, such as discomfort, excessive perspiration, pressure sores caused by non-uniform pressure distribution over the stump and the need for repeated fittings due to poor fit.1-3 Osseointegrated transcutaneous prostheses are able to avoid these problems because the forces that would be encountered by the soft tissues of the stump with the use of a socket prosthesis are transferred directly to the skeleton.

Achievement of a seal between the soft tissues and the implant is necessary to avoid infection. Soft-tissue attachment to the implant prevents epithelial downgrowth (the migration of epithelial cells downwards and parallel to the side of the implant), which may lead to marsupialisation (pocket-formation) of the soft tissues and thereby a route for organisms to enter the underlying soft tissues.4 In order to achieve soft-tissue integration, fibroblasts must win the ‘race for the surface’ against bacteria. The ‘race for the surface’ describes the competition between eukaryotic cell and bacterial adhesion on the implant’s surface.5,6 In the case of ITAP, fibroblasts need to adhere to the surface before bacteria. If fibroblasts win the ‘race’, the implant would be covered by a cellular layer, making the surface less available for bacterial attachment. This would lead to successful tissue integration. However, bacteria may win the ‘race’ establishing biofilm formation, which may be difficult to eradicate. The current ITAP design has a flange with drilled holes to increase soft-tissue integration. Infection, nevertheless, remains a risk.

This study aimed to investigate if surface modifications and coatings could improve soft-tissue integration by enabling fibroblasts to win the ‘race for the surface’ against bacteria. Fully porous titanium (i.e. with interconnected pores) has been shown to increase tissue integration, as pores allow migration and proliferation of cells, ingrowth of blood vessels and consequently tissue formation within the porous structure.7,8 Other osseointegrated prosthetic designs have shown improvements in soft-tissue integration with porous-coated implants, but have not eliminated the problem of marsupialisation.9 A fully porous ITAP flange has not previously been investigated. Fibronectin-functionalised hydroxyapatite (HAFn) has been shown to enhance in vitro fibroblast adhesion and, when applied to drilled-holed implants, has improved in vivo soft-tissue integration.10,11 However, it is known that bacteria often adhere to surfaces that promote soft-tissue integration and therefore, the presence of an antimicrobial coating during the early post-operative period before a robust seal has formed may be of value to prevent bacteria from adhering before fibroblasts attach. Silver has a broad spectrum of antimicrobial activity and can be incorporated into fibronectin-functionalised hydroxyapatite coatings using an electrochemical deposition technique.12,13 An advantage of the electrochemical deposition technique is that it may be used to coat complex structures (such as a porous flange). More commonly used line-of-sight techniques, such as plasma spraying, would not be able to do this.14 Fibronectin-functionalised hydroxyapatite coatings with silver (HAAgFn) have been shown to prevent bacterial colonisation while promoting viable fibroblast attachment in vitro.15

The hypothesis of this in vivo study is that a fully porous ITAP flange would increase soft-tissue integration over the current drilled-hole design and that HAAgFn coatings would further increase soft-tissue integration.

Materials and Methods

Transcutaneous pin designs and coatings

Transcutaneous pins with a laser-sintered porous titanium alloy flange (pore size 700 μm and strut size 300 μm, porosity 18%) (Eurocoatings, Trentino, Italy) were used (Fig. 1a, Table I). The flanges were either uncoated (PT) or coated with electrochemically deposited hydroxyapatite (PT-HA), hydroxyapatite (HA) with fibronectin (PT-HAFn), HA with silver (PT-HAAg) or HA with silver and fibronectin (PT-HAAgFn). The details of the surface preparation and electrochemical deposition process is available in the supplementary material. The current ITAP drilled-flange (DF) design, which consists of a flat flange with drilled holes (with a diameter of 700 μm) and a tapered intraosseous stem was used as a control (Fig. 1b, Table I). The flange and the stem are coated with plasma sprayed HA (Plasma Biotal Ltd, Buxton, United Kingdom). An uncoated straight pin (SP) without a flange was included as a control for the DF design.

Figs. 1a - 1b 
            Diagrams showing an Intraosseous
Amputation Prosthesis pin with a (a) porous titanium alloy flange
and (b) drilled flange.
Figs. 1a - 1b 
            Diagrams showing an Intraosseous
Amputation Prosthesis pin with a (a) porous titanium alloy flange
and (b) drilled flange.

Figs. 1a - 1b

Diagrams showing an Intraosseous Amputation Prosthesis pin with a (a) porous titanium alloy flange and (b) drilled flange.

Table I

Transcutaneous flanged-pin dimensions

Dimension (mm) PT DF
Total pin length 45 47
Length of transcutaneous portion 19.2 17
Length of intraosseous portion 21.5 29
Flange length 4.3 1
Pin diameter 5 5
Flange diameter 11 11
Thread pitch 0.8 -
Core diameter 4.2 -
  1. PT, porous titanium alloy flange; DF, drilled flange

Surface characterisation

Scanning electron microscopy (SEM) was performed to visualise the surface topography and morphology of all coatings on the outer surface of the implants and to measure the thickness of HA and HAAg coatings on the outer surface of the implant and within the central pores. Energy-dispersive X-ray spectroscopy analysis was performed to determine the Ca/P ratio of HA coatings and to measure the atomic percentage of Ag within coatings. A full description of this technique is available in the supplementary material.

Silver release

The amount of silver released into fetal calf serum from PT-HAAg was measured over a four-week period, with PT-HA used as a control. The flanges were placed in 5 mL of fetal calf serum at room temperature. Flanges were removed at four, eight and 24 hours and then on days 2, 3, 4, 5, 8, 11, 14, 18 and 28. Flanges were placed into fresh fetal calf serum solution at each time point. Inductively Coupled Plasma Mass Spectrometry (ICP-MS) (Trace Element Laboratory, London, United Kingdom) was used to quantify the silver content in elution fluids.

Animal model

The study was performed under a Project License and a Personal License according to the United Kingdom Home Office Scientific Procedures Act 1986. Skeletally mature female sheep (mules) (n = 6) weighing between 66 kg and 88 kg were used. Transcutaneous pins were implanted along the length of the sheep tibia. Details of the surgical procedure are included in the supplementary material.

Hard grade resin-histological processing

The specimens were fixed in 10% formal saline. Following this, they underwent ascending graded alcohol dehydration and defatting in chloroform. They were embedded in LR White Resin (London Resin Company Ltd, Reading, United Kingdom). The histological sectioning process is described in the supplementary material.

Hard grade resin-histological analysis

Two sections were analysed for each implant. The following outcomes were measured:

- epithelial downgrowth (mm)

- percentage epithelial attachment

- percentage dermal attachment.

A line intercept method was used for these measurements (Fig. 2).4 The thickness of the epithelial and dermal tissue layers was measured and expressed as a percentage of the thickness of these layers in contact with the implant surface.

Fig. 2 
            Schematic diagram showing the measurements
for epithelial downgrowth, epithelial and subepithelial attachment
for a flanged implant.

Fig. 2

Schematic diagram showing the measurements for epithelial downgrowth, epithelial and subepithelial attachment for a flanged implant.

A semi-quantitative estimate was used to express the percentage of the pore filled with soft tissue and the density of the tissue within the pore compared with that outside it. Fibroblast nuclei within the inner pores were counted within a defined area and the number per mm2 calculated. The blood vessels within the inner pores were counted within a defined area and number of blood vessels per mm2 was calculated.

Statistical analysis

This was performed using SPSS version 21.0 software (IBM, Armonk, New York). The Kruskal-Wallis test was used to identify differences between groups. Paired Mann-Whitney U tests were used to determine differences between implant types. A p-value < 0.05 was considered statistically significant. Data are presented as median values and 95% confidence intervals (CI).

Results

Surface characterisation-surface topography, coating thickness and elemental composition

The electrochemically deposited coatings on the porous titanium alloy flanges contained a combination of plate-like, needle-shaped and globular crystals similar to that observed on solid discs in previous in vitro studies.15 SEM images are shown in Figure 3.

Figs. 3a - 3b 
            Scanning electron micrographs of
(a) the outer surface of a hydroxyapatite (HA) with silver and fibronectin-coated
implant (× 50 magnification) and (b) a section through the centre
of a hydroxyapatite coated implant (PT) showing the coating on the
innermost pores.
Figs. 3a - 3b 
            Scanning electron micrographs of
(a) the outer surface of a hydroxyapatite (HA) with silver and fibronectin-coated
implant (× 50 magnification) and (b) a section through the centre
of a hydroxyapatite coated implant (PT) showing the coating on the
innermost pores.

Figs. 3a - 3b

Scanning electron micrographs of (a) the outer surface of a hydroxyapatite (HA) with silver and fibronectin-coated implant (× 50 magnification) and (b) a section through the centre of a hydroxyapatite coated implant (PT) showing the coating on the innermost pores.

The median calcium:phosphate atomic percentage ratio for electrochemically deposited HA coatings was 1.74 (95% CI 1.62 to 1.86). This indicated that these HA coatings were close in composition to pure stoichiometric HA. The thickness of coatings is presented in Table II. Table III shows the silver atomic percentages of HAAg/Fn coatings.

Table II

Thickness of coatings

Surface Coating thickness (μm) median value (95% confidence interval)
PT-HA (outer surface) 47 (30 to 76)
PT-HA (inner pores) 36 (12 to 55)
PT-HAAg (outer surface) 56 (24 to 101)
PT-HAAg (inner pores) 46 (31 to 82)
  1. PT-HA, hydroxyapatite coated flange; PT-HAAg, hydroxyapatite with silver coated flange

Table III

Silver percentages within PT-HAAg and PT-HAAgFn coatings measured by energy-dispersive X-ray spectroscopy analysis

Surface Ag atomic % median value (95% confidence interval)
PT-HAAg 0.700 (0.582 to 0.757)
PT-HAAgFn 0.610 (0.591 to 0.652)
  1. PT-HAAg, hydroxyapatite with silver coated flange; PT-HAAgFn, hydroxyapatite with silver and fibronectin coated flange

Silver release

The silver release was greatest over the first 24 hours, and gradually reduced over time (4.14 μmol/L at 28 days). The rate of release was slowest between days 11 and 28. Silver continued to be released throughout the 28-day test period (Table IV).

Table IV

The release of silver from hydroxyapatite with silver coated flange over 28 days

Time (days) Silver release (umol/L) median value (95% confidence interval)
1 101.5810 (46.7192 to 163.9788)
2 27.6840 (0.5311 to 60.5682)
3 21.9790 (3.5346 to 43.1121)
4 23.5450 (3.5155 to 40.7465)
5 19.4430 (6.0237 to 31.2776)
8 12.1270 (-1.5545 to 27.2212)
11 5.3860 (2.6626 to 7.5661)
14 4.3950 (1.4158 to 6.5149)
18 3.5550 (0.6460 to 5.4300)
28 4.1400 (-1.0457 to 7.7223)

Soft-tissue integration-epithelial downgrowth

SP resulted in the greatest degree of epidermal downgrowth. DF reduced epithelial downgrowth compared with SP (DF = 4.2 mm, SP = 6.1 mm; Mann-Whitney U test, p < 0.001). PT reduced epithelial downgrowth compared with DF (Mann-Whitney U test, p = 0.027). The addition of coatings to PT did not result in any further reduction in epidermal downgrowth. PT-HAFn, PT-HAAg and PT-HAAgFn reduced epidermal downgrowth compared with DF (Mann-Whitney U test, p = 0.003, p = 0.017 and p = 0.001 respectively). The reduction in epidermal downgrowth conferred by the PT-HA coating compared with DF did not reach statistical significance (Mann-Whitney U test, p = 0.052) (Table V).

Table V

Epithelial downgrowth measurements associated with each implant

Surface Epithelial downgrowth (mm) median value (95% confidence interval)
Straight pin 6.0753 (5.2082 to 8.7915)
DF 4.2043 (2.6040 to 4.3087)
PT 1.2900 (0.9936 to 3.1139)
PT-HA 1.7826 (1.2789 to 3.3195)
PT-HAFn 1.7655 (1.2074 to 2.3601)
PT-HAAg 1.7562 (0.9242 to 2.9181)
PT-HAAgFn 1.7888 (1.3366 to 2.1510)
  1. DF, drilled flange; PT, uncoated porous titanium alloy flange; PT-HA, hydroxyapatite coated flange; PT-HAFn, hydroxyapatite with fibronectin coated flange; PT-HAAg, hydroxyapatite with silver coated flange; PT-HAAgFn, hydroxyapatite with silver and fibronectin coated flange

Soft-tissue integration-percentage epithelial attachment

The epithelium did not attach to the SP and this implant was therefore associated with the lowest median percentage epithelial attachment (0%, interquartile range (IQR) 0% to 0%). The DF had a greater median percentage epithelial attachment, which was highly variable between implants (63.4%, IQR 0% to 100%) (Mann-Whitney U test, p = 0.045). There was no difference between DF and PT implants, irrespective of coating (Kruskal-Wallis test, p = 0.859).

Soft-tissue integration-percentage dermal attachment

The DF showed greater median dermal attachment than the SP (DF = 56.4%, SP = 0%; Mann-Whitney U test, p = 0.017). The PT component had greater median dermal attachment than DF (PT = 80.5%; Mann-Whitney U test, p = 0.020), with the addition of coatings conferring no statistically significant improvements in dermal attachment. Improved attachment was seen in PT-HAFn compared with PT-HA (Mann-Whitney U test, p = 0.023) and in PT-HAAgFn compared with PT-HAAg (Mann-Whitney U test, p = 0.039) (Fig. 4). Histological sections are shown in Figure 5.

Fig. 4 
            Box-and-whisker plot showing the percentage
dermal attachment to implants (SP, straight pin; DF, drilled flange; PT,
uncoated porous titanium alloy flange; PT-HA, hydroxyapatite coated
flange; PT-HAFn, hydroxyapatite with fibronectin coated flange;
PT-HAAg, hydroxyapatite with silver coated flange; PT-HAAgFn, hydroxyapatite
with silver and fibronectin coated flange).

Fig. 4

Box-and-whisker plot showing the percentage dermal attachment to implants (SP, straight pin; DF, drilled flange; PT, uncoated porous titanium alloy flange; PT-HA, hydroxyapatite coated flange; PT-HAFn, hydroxyapatite with fibronectin coated flange; PT-HAAg, hydroxyapatite with silver coated flange; PT-HAAgFn, hydroxyapatite with silver and fibronectin coated flange).

Figs. 5a - 5d 
            Histological sections showing
(a) incomplete soft-tissue fill within a drilled flange (DF) pore
(P); (b) areas of lack of the dermal tissue (D) attachment to the
flange of a DF implant (arrows); (c) intimate contact between uncoated porous
titanium alloy flange (PT) pore edges and soft-tissue PT. Increased
collagen deposition is seen at the pore periphery and (d) the dermis
attaching to hydroxyapatite with fibronectin coated flange flange.
Figs. 5a - 5d 
            Histological sections showing
(a) incomplete soft-tissue fill within a drilled flange (DF) pore
(P); (b) areas of lack of the dermal tissue (D) attachment to the
flange of a DF implant (arrows); (c) intimate contact between uncoated porous
titanium alloy flange (PT) pore edges and soft-tissue PT. Increased
collagen deposition is seen at the pore periphery and (d) the dermis
attaching to hydroxyapatite with fibronectin coated flange flange.
Figs. 5a - 5d 
            Histological sections showing
(a) incomplete soft-tissue fill within a drilled flange (DF) pore
(P); (b) areas of lack of the dermal tissue (D) attachment to the
flange of a DF implant (arrows); (c) intimate contact between uncoated porous
titanium alloy flange (PT) pore edges and soft-tissue PT. Increased
collagen deposition is seen at the pore periphery and (d) the dermis
attaching to hydroxyapatite with fibronectin coated flange flange.
Figs. 5a - 5d 
            Histological sections showing
(a) incomplete soft-tissue fill within a drilled flange (DF) pore
(P); (b) areas of lack of the dermal tissue (D) attachment to the
flange of a DF implant (arrows); (c) intimate contact between uncoated porous
titanium alloy flange (PT) pore edges and soft-tissue PT. Increased
collagen deposition is seen at the pore periphery and (d) the dermis
attaching to hydroxyapatite with fibronectin coated flange flange.

Figs. 5a - 5d

Histological sections showing (a) incomplete soft-tissue fill within a drilled flange (DF) pore (P); (b) areas of lack of the dermal tissue (D) attachment to the flange of a DF implant (arrows); (c) intimate contact between uncoated porous titanium alloy flange (PT) pore edges and soft-tissue PT. Increased collagen deposition is seen at the pore periphery and (d) the dermis attaching to hydroxyapatite with fibronectin coated flange flange.

Soft-tissue integration-percentage soft-tissue fill

The median percentage soft-tissue fill within the inner pores of PT was significantly greater than that of the DF (PT = 90%, DF = 65%; Mann-Whitney U test, p < 0.001). The addition of coatings did not result in any increases in percentage soft-tissue fill (Table VI). Histological sections are shown in Figure 5.

Table VI

Number of cell nuclei/mm2 within the inner pores of each implant

Surface Cell nuclei/mm2median value (n, 95% confidence interval)
DF 1520.5050 (1028.8516 to 2332.6951)
PT 3922.6250 (2590.6413 to 5182.6021)
PT-HA 2883.8250 (2029.4570 to 3891.7696)
PT-HAFn 2945.8300 (2473.5655 to 3708.2709)
PT-HAAg 3086.2450 (2595.8449 to 4017.1911)
PT-HAAgFn 2606.3700 (1990.1559 to 4334.6132)
  1. DF, drilled flange; PT, uncoated porous titanium alloy flange; PT-HA, hydroxyapatite coated flange; PT-HAFn, hydroxyapatite with fibronectin coated flange; PT-HAAg, hydroxyapatite with silver coated flange; PT-HAAgFn, hydroxyapatite with silver and fibronectin coated flange

Soft-tissue integration-number of cell nuclei/mm2

There was a greater median density of cell nuclei within the inner pores of the PT flange than the DF (PT = 3922.6 mm2, DF = 1520.5 mm2; Mann-Whitney U test, p = 0.018). The addition of coatings was not associated with any change in density (Table VII).

Table VII

Number of cell nuclei/mm2 within the inner pores of each implant

Surface Cell nuclei/mm2median value (n, 95% confidence interval)
DF 1520.5050 (1028.8516 to 2332.6951)
PT 3922.6250 (2590.6413 to 5182.6021)
PT-HA 2883.8250 (2029.4570 to 3891.7696)
PT-HAFn 2945.8300 (2473.5655 to 3708.2709)
PT-HAAg 3086.2450 (2595.8449 to 4017.1911)
PT-HAAgFn 2606.3700 (1990.1559 to 4334.6132)
  1. DF, drilled flange; PT, uncoated porous titanium alloy flange; PT-HA, hydroxyapatite coated flange; PT-HAFn, hydroxyapatite with fibronectin coated flange; PT-HAAg, hydroxyapatite with silver coated flange; PT-HAAgFn, hydroxyapatite with silver and fibronectin coated flange

Soft-tissue integration-number of blood vessels/mm2

There was a greater median density of blood vessels within the inner pores of the PT flange than the DF (PT = 213.2 mm2, DF = 0 mm2; Mann-Whitney U test, p = 0.002). Coatings did not result in further increases in blood vessel ingrowth.

Discussion

This study has shown for the first time that using a PT flange reduces epithelial downgrowth and increases dermal attachment and ingrowth of vascularised soft tissue into pores when compared with the standard drilled-hole ITAP flange. Other studies of the effect of porous titanium alloy on soft-tissue integration of transcutaneous pins have used porous coatings rather than a fully porous material, have assessed smaller pore sizes and have made comparisons with smooth implants rather than drilled-hole implants.9,16,17

The coatings did not show any statistically significant advantages over PT. The median dermal attachment of PT was 80% and the median soft-tissue fill was 90%. It may be difficult to achieve significant improvements or complete soft-tissue attachment with the animal model used in this study. In this study, the sheep were mobilised bearing full weight from the first post-operative day, which would create some movement at the skin-implant interface before the seal could form. In clinical practice in humans, there is a rest period where the implant is not loaded during the early wound-healing period. Increased bacterial exposure is also likely in an animal model, however, the sheep may have a higher resistance to bacterial infections than humans.

Although there were no statistically significant differences when coatings were added to PT, PT-HA lost some of the beneficial effects associated with PT. The reduction in epithelial downgrowth associated with PT-HA compared with DF was not statistically significant. PT with all other coatings significantly reduced epithelial downgrowth. Additionally, PT-HA was the only PT implant not associated with significantly improved blood vessel ingrowth compared with DF. The negative effect of HA may be due to increased bacterial colonisation when this coating is used on porous titanium, permitting bacteria to win the ‘race for the surface’.15 PT-HAFn did not have a negative effect although, similarly to HA, HAFn would be expected to promote adhesion of some bacterial strains.18-20 It is possible that when fibronectin is added to HA, despite its potential to increase bacterial colonisation, cells are able to win the ‘race for the surface’ due to early soft-tissue adherence allowing HAFn to overcome any loss of beneficial effects associated with HA.

Silver release was sustained over the 28 days at a slow rate indicating that the coating persists for 28 days. These were below levels that would cause systemic toxicity.21 The levels released remained above those shown to have bactericidal activity in vitro.12 HAAg coatings did not produce any adverse soft-tissue reactions suggesting that the levels of silver were safe. PT-HAAg was associated with increased vascularisation within the inner pores compared with PT-HA. Additionally, PT-HAAg/Fn maintained the beneficial effects that PT had on epithelial downgrowth and vascularity over DF that PT-HA had lost. This indicates that silver had some effect. The effect of silver coatings without HA or increasing the concentration of silver may have a different effect. It is possible that, despite not having a significant effect on soft-tissue attachment, silver could reduce bacterial colonisation, which could be of value in terms of preventing infection. If this were to be the case, the fact that this study has shown that there is no adverse soft-tissue reaction associated with silver, would be useful. Further data on bacterial colonisation on the implant surfaces in this study will be published separately.

It has been previously shown that if dermal attachment is achieved even without epithelial attachment, epithelial downgrowth is inhibited.4 Once epithelial downgrowth is prevented and keratinocytes in the epithelium come into contact with the implant surface, they may attach via hemidesmosomes and allow an epithelial seal to form in addition to the dermal seal.4,22,23 The porous flanges did not, however, significantly increase epithelial attachment, despite the increase in dermal attachment. Design modifications to increase epithelial attachment may be necessary, such as coating the transcutaneous portion of the implant with proteins that enhance keratinocyte attachment.23

There are several potential limitations of this study. Firstly, we measured soft-tissue contact rather than mechanical strength of attachment. However, other studies that have performed mechanical testing have shown that the histological findings support strength of attachment results.24,25 Secondly, the time period of four weeks for this study was short. It is known that as time increases soft-tissue attachment increases.24,26 The short time period which was selected as an assessment of early attachment is essential, as early soft-tissue attachment is the key factor affecting the outcome of the ‘race for the surface’.5,6 Longer periods of time would also make quantification of cell nuclei density challenging due to the proliferation of nuclei.

Although the main intended application of the findings from this research is for improving the success of ITAP, the findings could also be applied to other transcutaneous devices. For example, the addition of a porous flange to external fixator pins may have the potential to reduce infection rates by improving soft-tissue integration at the skin–implant interface. Furthermore, the coating HAAgFn may be more likely to produce favourable in vivo results for osseointegration than for soft-tissue integration due to the rough topography. Further investigation of porous metals coated with electrochemically deposited HAAg (with and without fibronectin) to assess the effect on osseointegration would be useful to determine if these coatings could play a role in preventing loosening and bacterial colonisation of the intraosseous portion of ITAP or for uncemented arthroplasty surgery. Additionally, this coating technique could be applied to endoprosthetic replacements with porous collars e.g. for tumour resection.

In conclusion, the results of this study indicate that a PT flange, without the addition of coatings, has the potential to increase the soft-tissue seal around ITAP, which would be expected to reduce the susceptibility of ITAP to infection compared with the current drilled-hole flange design. Further investigation of fibronectin and silver coatings without HA may be of value.

Take home message:

- A fully porous titanium alloy flange with interconnected pores improves soft-tissue integration. It reduces epithelial downgrowth and increases dermal attachment, cell nuclei density and blood vessel ingrowth compared with the current standard ITAP model.

- The addition of hydroxyapatite, silver and fibronectin coatings did not significantly improve soft-tissue integration.

- It is therefore postulated that a fully porous titanium alloy flange, without the addition of coatings, has the potential to reduce the susceptibility of ITAP to infection by increasing soft-tissue integration.


Correspondence should be sent to M. Chimutengwende-Gordon; email:

1 Hagberg K , BrånemarkR. Consequences of non-vascular trans-femoral amputation: a survey of quality of life, prosthetic use and problems. Prosthet Orthot Int2001;25:186194.CrossrefPubMed Google Scholar

2 Koc E , TuncaM, AkarA, et al.Skin problems in amputees: a descriptive study. Int J Dermatol2008;47:463466.CrossrefPubMed Google Scholar

3 Meulenbelt HE , GeertzenJH, JonkmanMF, DijkstraPU. Skin problems of the stump in lower limb amputees: 1. A clinical study. Acta Derm Venereol2011;91:173177.CrossrefPubMed Google Scholar

4 Pendegrass CJ , GoodshipAE, BlunnGW. Development of a soft tissue seal around bone-anchored transcutaneous amputation prostheses. Biomaterials2006;27:41834191.CrossrefPubMed Google Scholar

5 Gristina AG . Biomaterial-centered infection: microbial adhesion versus tissue integration. Science1987;237:15881595.PubMed Google Scholar

6 Subbiahdoss G , PidhatikaB, CoullerezG, et al.Bacterial biofilm formation versus mammalian cell growth on titanium-based mono- and bi-functional coating. Eur Cell Mater2010;19:205213.CrossrefPubMed Google Scholar

7 Hulbert SF , YoungFA, MathewsRS, et al.Potential of ceramic materials as permanently implantable skeletal prostheses. J Biomed Mater Res1970;4:433456.CrossrefPubMed Google Scholar

8 Kuboki Y , TakitaH, KobayashiD, et al.BMP-induced osteogenesis on the surface of hydroxyapatite with geometrically feasible and nonfeasible structures: topology of osteogenesis. J Biomed Mater Res1998;39:190199.CrossrefPubMed Google Scholar

9 Jeyapalina S , BeckJP, BachusKN, WilliamsDL, BloebaumRD. Efficacy of a porous-structured titanium subdermal barrier for preventing infection in percutaneous osseointegrated prostheses. J Orthop Res2012;30:13041311.CrossrefPubMed Google Scholar

10 Pendegrass CJ , MiddletonC, BlunnW. Fibronectin Functionalized Hydroxyapatite Coatings: Improving Dermal Fibroblast Adhesion In Vitro and In Vivo. Adv Eng Mater2010;12:B365B373. Google Scholar

11 Pendegrass CJ , El-HusseinyM, BlunnGW. The development of fibronectin-functionalised hydroxyapatite coatings to improve dermal fibroblast attachment in vitro. J Bone Joint Surg [Br]2012;94-B:564569.CrossrefPubMed Google Scholar

12 Ghani Y , CoathupMJ, HingKA, BlunnGW. Development of a hydroxyapatite coating containing silver for the prevention of peri-prosthetic infection. J Orthop Res2012;30:356363.CrossrefPubMed Google Scholar

13 Brennan SA , Ní FhoghlúC, DevittBM, et al.Silver nanoparticles and their orthopaedic applications. Bone Joint J2015;97-B:582589.CrossrefPubMed Google Scholar

14 García-Gareta E , HuaJ, KnowlesJC, BlunnGW. Comparison of mesenchymal stem cell proliferation and differentiation between biomimetic and electrochemical coatings on different topographic surfaces. J Mater Sci Mater Med2013;24:199210.CrossrefPubMed Google Scholar

15 Chimutengwende-Gordon M , PendegrassC, BaystonR, BlunnG. Preventing infection of osseointegrated transcutaneous implants: incorporation of silver into preconditioned fibronectin-functionalized hydroxyapatite coatings suppresses Staphylococcus aureus colonization while promoting viable fibroblast growth in vitro. Biointerphases2014;9:031010.CrossrefPubMed Google Scholar

16 Isackson D , McGillLD, BachusKN. Percutaneous implants with porous titanium dermal barriers: an in vivo evaluation of infection risk. Med Eng Phys2011;33:418426.CrossrefPubMed Google Scholar

17 Farrell BJ , PrilutskyBI, RitterJM, et al.Effects of pore size, implantation time, and nano-surface properties on rat skin ingrowth into percutaneous porous titanium implants. J Biomed Mater Res A2014;102:13051315.CrossrefPubMed Google Scholar

18 Dexter SJ , PearsonRG, DaviesMC, CámaraM, ShakesheffKM. A comparison of the adhesion of mammalian cells and Staphylococcus epidermidis on fibronectin-modified polymer surfaces. J Biomed Mater Res2001;56:222227.CrossrefPubMed Google Scholar

19 Plotkowski MC , FilhoMB, de Nazareth MeirellesM, TournierJM, PuchelleE. Pseudomonas aeruginosa Binds to Soluble Cellular Fibronectin. Curr Microbiol1993;26:9195. Google Scholar

20 Bozzini S , VisaiL, PignattiP, PetersenTE, SpezialeP. Multiple binding sites in fibronectin and the staphylococcal fibronectin receptor. Eur J Biochem1992;207:327333.CrossrefPubMed Google Scholar

21 Perrelli G , PiolattoG. Tentative reference values for gold, silver and platinum: literature data analysis. Sci Total Environ1992;120:9396.CrossrefPubMed Google Scholar

22 Pendegrass CJ , GoodshipAE, PriceJS, BlunnGW. Nature’s answer to breaching the skin barrier: an innovative development for amputees. J Anat2006;209:5967. Google Scholar

23 Gordon DJ , BhagawatiDD, PendegrassCJ, MiddletonCA, BlunnGW. Modification of titanium alloy surfaces for percutaneous implants by covalently attaching laminin. J Biomed Mater Res A2010;94:586593.CrossrefPubMed Google Scholar

24 LaBerge M , BobynJD, RivardCH, DrouinG, DuvalP. Study of soft tissue ingrowth into canine porous coated femoral implants designed for osteosarcomas management. J Biomed Mater Res1990;24:959971.CrossrefPubMed Google Scholar

25 Hacking SA , BobynJD, TohK, TanzerM, KrygierJJ. Fibrous tissue ingrowth and attachment to porous tantalum. J Biomed Mater Res2000;52:631638.CrossrefPubMed Google Scholar

26 Bobyn JD , WilsonGJ, MacGregorDC, PilliarRM, WeatherlyGC. Effect of pore size on the peel strength of attachment of fibrous tissue to porous-surfaced implants. J Biomed Mater Res1982;16:571584.CrossrefPubMed Google Scholar

Author contributions:

M. Chimutengwende-Gordon: Study design, Surgery, Histology, Data collection and analysis, Writing the paper.

C. Pendegrass: Supervision, Data/manuscript review.

G. Blunn: Supervision, Data/manuscript review.

This research was funded by an MRC Clinical Research Training Fellowship awarded to M. Chimutengwende-Gordon.

No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.

This is an open-access article distributed under the terms of the Creative Commons (CC-BY) license, which permits copying and redistributing the material in any medium or format, remixing, transforming and building upon the material for any purpose, even commercially, provided the original author and source are credited, and changes made are indicated. This may be done in a reasonable manner, but not in any way that suggests the licensor endorses you or your use.

This article was primary edited by P. Page and first proof edited by G. Scott.

Supplementary material. Further information regarding the surface preparation, the surgical details and the histological section is included alongside the online version of this article at www.bjj.boneandjoint.org.uk.