header advert
Orthopaedic Proceedings Logo

Receive monthly Table of Contents alerts from Orthopaedic Proceedings

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Visit Orthopaedic Proceedings at:

Loading...

Loading...

Full Access

115 – MINIMUM TEN-YEAR RESULTS OF A DUAL OFFSET UNCEMENTED TAPERED FEMORAL STEM FOR THR



Abstract

Purpose: The Synergy femoral component was introduced in late 1996 as a second generation titanium proximally porous-coated tapered stem with dual offsets (standard & high) to help better restore femoral offset at THR. The purpose of this prospective study was to evaluate the long-term (minimum 10 year) clinical and radiographic results and survivorship of this second-generation femoral component at our institution.

Method: From December 1996 to December 1999, 256 cementless Synergy femoral components were inserted and followed prospectively in 254 patients requiring THR. 185 were standard offset stems(72.3%) while 71 stems (27.7%) were high offset. The average follow-up was 11.2 years (range 10.0 to 12.9 years). Average age at index THR was 58.9 years (range 19 to 86 years). Two hundred eight stems had standard porous coating while 48 had additional HA coating. Fifty-two cases were either lost to follow-up or had died prior to 10 years follow-up. Patients were followed prospectively using validated clinical outcome scores (WOMAC, SF-12, Harris Hip scores) and radiographs. Kaplan-Meier survival analysis was performed.

Results: All health-related outcomes were significantly improved from pre-operative with a mean Harris Hip score and WOMAC at last follow-up of 91.6 and 81.8, respectively. From the initial 256 femoral stems inserted, only 5 stems have been revised. Two stems have been revised for infection. To date, only one stem has been revised due to subsidence at nine months following surgery likely, as a result of a calcar fracture occurring at the index THR. Two stems were revised for peri-prosthetic fracture as a result of patient falls at six months and 9.8 years post-op. Radiographic review of remaining stems in-situ identified no cases of loosening with all stems showing evidence of osseous integration. The Kaplan-Meier survivorship analysis of the femoral component, with revision for aseptic loosening, was 99.2%±0.008 at five, 10 and 12 years.

Conclusion: The Synergy femoral component, a second generation titanium proximally porous-coated tapered stem design with dual offsets, has demonstrated excellent clinical & radiographic results and long-term survivorship (99% at 10 years) at our institution.

Correspondence should be addressed to: COA, 4150 Ste. Catherine St. West Suite 360, Westmount, QC H3Z 2Y5, Canada. Email: meetings@canorth.org